Transmitter mag-flux M1
Connection of the sensor
mag-flux
S and the flow probes
mag-flux
MIS
These sensors are equipped with a pre-assembled cable ex-
factory. This cable is permanently connected to the sensor at one
end. The end leading to the transmitter is fitted with a cable
gland and pre-prepared cable ends for connection. The attached
wire numbers serve for orientation.
Fig 18
Connections of the sensor
mag-flux
S and the probes
Anschluss HART
®
A number of options are available for HART
®
communication.
However, for all these options loop resistance must be less than
the maximum load specified in Chapter „Outputs“ (see page 6).
The HART
®
-Interface is connected via terminals 8 and 9 of the
(active) current output.
The minimum load impedance is 250Ω.
.
Fig 19
Electrical connection for HART
®
communication, schematic diagram
Display and operator interface (basic version)
Zero point adjustment
Below the mains transformer, next to the lights and the data
storage module (DSB), there is a switch for adjusting the zero
point. In order to reach the switch, the transducer must be
opened by unscrewing the front cover and removing the
decoration foil.
For processing the zero point adjustment please observe the
advices in Section “Zero point calibrationon” on page 4.
Fig 20
Switch for zero point adjustment
LED display
The actual operating status information of the
mag-flux M1
is
displayed by three LED for the basic version:
Fig 21
Status-LED of the
mag-flux M1
(basic version)
off
Not powered
green LED
flashing
Power on
off
Flow direction forward
yellow LED
on
Flow direction reverse
off
Normal operation
flashing
Limit exceeded
red LED
on
Malfunktion