Meco 27-AUTO Instruction Manual Download Page 3

225AUTO : 28-09-2012

AUTORANGING

DIGITAL CLAMP METER

MODEL : 

225

AUTO

INSTRUCTION  MANUAL

®

1.  SPECIFICATIONS

1.1 General Specification
Display

n

3

¾

 digit liquid crystal display (LCD) with a

maximum reading of 3999.

Polarity

n

Automatic, positive implied, negative polarity
indication.

Overrange

n

(OL) or  (-OL)  is displayed.

Zero

n

Automatic.

Low battery

n

“       ” is displayed when the battery

indication

voltage drops below the operating level.

Measurement

n

3 times per second, nominal.

rate
Operating

n

0°C to 50°C at < 70% relative humidity.

Environment
Storage

n

-20°C to 60°C, 0 to 80% R.H. with battery

Temperature

removed from meter.

Accuracy

n

Stated accuracy at 27

0

C ± 5

0

C,

<75% relative humidity.

Power

n

Two  1.5V  ‘AAA’  Size Battery

Battery life

n

200 hours typical

Dimensions

n

186 x 58 x 30mm (approx.)

Weight

n

165gms including battery (approx.)

Accessories

n

Pair of test leads x 1, instruction manual x 1,
1.5V batterys (installed) x 2, Carring Case x 1

Maximum Jaw

n

32mm

Opening

1.2 Electrical Specification

Accuracies are ± (% r number of digits) at 27 ± 5°C and
humidity of less than 75% RH.

AC CURRENT (50-60Hz)

DC VOLTAGE (Auto Ranging)

Range

Accuracy

Overload Protection

40A

±(1.5% rdg + 5dgt)

400A AC Max.

400A

for 1 minute

Range

Accuracy

Overload Protection

600V

±(0.8% rdg + 1dgt)

AC voltage ranges 600V AC

Resistance ranges 250V AC

Range

Accuracy

Overload Protection

600V

±(1.0% rdg + 3dgt)

AC voltage ranges 600V AC

Resistance ranges 250V AC

RESISTANCE (Auto Ranging)

Range

Accuracy

Overload Protection

40M

V

V

V

V

V

±(1.0% rdg + 2dgt)

AC voltage ranges 600V AC

Resistance ranges 250V AC

AC VOLTAGE (Auto Ranging) 50-60Hz

Continuity Check

Threshold Level

: 50

V

 Approx.

Response Time

: 1m Sec. Approx.

Open Circuit Voltage: 0.4V Approx.
Indication

: ‘   ’ is displayed on LCD and buzzer sounds

at continuity.

Diode Test

Measurement Current   : 1.0 ± 0.6 mA Approx.
Open Circuit Voltage     : 0.4V Approx.

2. OPERATION

Before taking any measurements, read the safety information section.
Always examine  the instrument for damage, contamination (excessive
dirt,  grease, etc.) and defects. Examine the test leads for cracked or
frayed insulation. If any abnormal conditions exist do not attempt to
make any measurements.

SELECT BUTTON

The Button is used for select 

V

 / 

 /    range.

DATA HOLD BUTTON

Press Data Hold button to toggle in and out of Data Hold mode, In the
Data  Hold mode, the “ HOLD ” annunciator is displayed.

3

4

2

1

Reviews: