There are two external control ports at the rear of the VersaTest. The
Display Unit interface (b, in Fig.2) enables the VersaTest to respond, via
the appropriate Mecmesin cable, to signals from an AFG.  Please refer to
your AFG Operating Manual for further details.  Communication between
the VersaTest and other external devices (e.g. a PC) is possible via the
RS232 interface (c, in Fig.2) and an appropriate cable. 

The VersaTest must be powered off when attaching either cable.
When the interfaces are not in use please ensure that they are

covered with the supplied sheilds at all times.
For further information (on programming/protocol) contact Mecmesin
or their Approved Agent and ask for Technical Datasheet M/512051. 

MODEL

VERSATEST

Load capacity:

2500N (250kg, 550lb)

Limit switch repeatability:

<0.1mm* (0.004in)

Power consumption:

150 watts (maximum)

Over-run at top speed:

<2mm (0.008in)

Weight (stand only):

27kg (59.6lb)

Operating modes:

Manual, cyclic and single cycle

CROSSHEAD MOTION

Reverse on alarm point:

Yes, with AFG or AFTI, and cable

Travel range:

500mm (19.7in)

Reverse on sample break:

Yes, with AFG or AFTI, and cable

Maximum daylight:

620mm (24.4in)

ENVIRONMENTAL OPERATING CONDITIONS:

Max headroom (as in photograph)

510mm (20.08in)

Temperature range:

5

O

C - 40

O

C

Anvil plate:

160mm (6.3in)

Humidity:

<92.5%

x 100mm (3.94in)

STANDARD LOAD MEASUREMENT OPTIONS:

Female threads 10/32” UNF

Force gauge & dovetail bracket

5/16” UNC

S-beam loadcell, tension block module & AFTI

Speed range:

12.5 - 500mm/min

SPECIAL OPTIONS (details available on request)

(0.5 - 19.7in/min)

Increased crosshead travel:

Yes

Speed accuracy:

± 5% of indicated speed

Increased crosshead depth:

Yes, (load capacity is reduced)

Up and down settings:

Yes

Machine guard:

Yes

Speed indicated on stand:

Yes

Horizontal operation:

Yes

Direction of travel indicated on stand:

Yes

Simple logging/plotting PC software:

DataPlot (see M/501201)

Full computer control:

PCM (see M/512071)

* At given speed (i.e and not necessarily over whole speed range)

Communicating with External Devices

Communicating with External Devices

+

+

Fig 2

a

ON/OFF
Switch

b

Display Unit
Interface

c

RS232
Interface

VersaTest Technical Specification

VersaTest Technical Specification

4

4

Summary of Contents for VersaTest

Page 1: ...It is essential that you familiarise yourself with the contents of these instructions before attempting to operate the VersaTest English S I Instruments 256 South Rd Hilton South Australia 5033 Ph 08 8352 5511 info si instruments com www si instruments com ...

Page 2: ...aTest The dimensions speed range and capacity have been carefully optimised to cover the test procedures and specimen sizes most often needed by quality departments within many sectors of manufacturing industry The VersaTest has been designed and manufactured in a controlled system to ensure compliance with all relevant European Community Directives See Appendix 3 Additional information on the Ver...

Page 3: ...d the LED indicating direction of travel will illuminate The crosshead will cycle continuously between the limit switch positions During the cycle setting the Mode switch to Manual Control or selecting either the Up or Down position will stop the crosshead moving You will need to attach a force measuring device to your VersaTest this will usually be an Advanced Force Gauge or an S beam loadcell At...

Page 4: ...larm point Yes with AFG or AFTI and cable Travel range 500mm 19 7in Reverse on sample break Yes with AFG or AFTI and cable Maximum daylight 620mm 24 4in ENVIRONMENTAL OPERATING CONDITIONS Max headroom as in photograph 510mm 20 08in Temperature range 5O C 40O C Anvil plate 160mm 6 3in Humidity 92 5 x 100mm 3 94in STANDARD LOAD MEASUREMENT OPTIONS Female threads 10 32 UNF Force gauge dovetail bracke...

Page 5: ... as quickly as possible Installing the Machine 3 After placing the machine on a stable and level work surface check that the mains inlet voltage corresponds to your electrical installation either 230 Vac or 110 Vac The machine has a label close to the mains inlet connector which advises which voltage it is set for If the machine does not correspond to your supply inform your local Mecmesin supplie...

Page 6: ...ried out Machine Guarding 7 If after the assessment it is considered that machine guarding is needed then contact your local Mecmesin supplier who though the Mecmesin Application Engineering Department can arrange the supply of a suitable guard to offer the required level of protection Continued Safe Use 8 Once the machine is installed it should provide a reliable long term resource for universal ...

Page 7: ......

Reviews: