MEATEST, s.r.o.
M-142 Multifunction Calibrator
User Manual v48
95
[SOUR] :TEMP :PRT :NRESistance (?) <DNPD>
This command sets the nominal resistance of the resistance temperature sensor at 0
C. 10
to 2k
can be set.
<DNPD>
Represents the nominal resistance in Ohms.
If query is sent, M142 returns the set value of nominal resistance using standard exponential format. Example:
20.5
is returned as 2.001.
MEASure subsystem
This subsystem allows to control the internal multimeter of M142 calibrator. It sets the multimeter’s functions
and reads measured values.
Keyword
Parameters
MEASure
?
: CONFigure
: VOLTage
: CURRent
: MVOLTage
: RESistance
: FREQuency
: TEMPerature
: RTD
: TYPE (?)
<CPD> { PT385 | PT392 }
: NRESistance (?)
<DNPD>
: THERmocouple
: TYPE (?)
<CPD> { B | E | J | K | N | R | S | T }
: RJUNction (?)
<DNPD>
MEAS ?
This command returns the measured value.
Example:20.5
is returned as 2.001.
MEAS :CONF ?
M142 returns the set measurement mode { VOLT | CURR | MVOLT | RES | FREQ | TEMPerature:RTD |
TEMPerature:THERmocouple | OFF }.