This chapter explains how to perform the routine maintenance to keep your device in optimal operating
conditions.
9010+ calibrator has 2 user-replaceable fuses, both located on rear panel. Replace the fuses as follows:
1.
Switch the calibrator off and disconnect power cord from it
2.
Locate the fuse to be replaced: Mains fuse is in fuse/voltage selector case next to mains
connector, current output fuse is in the middle of the rear panel.
3.
Use flat screwdriver to either open voltage selector cover or turn current output fuse cover to
get to the fuse holder.
4.
Remove the fuse, check if it’s blown and replace it with new fuse of the same rating if needed.
To keep the device in mint condition, clean the case and front panel keys using a soft cloth slightly
dampened with either water or a non-abrasive mild cleaning solution that is not harmful to plastics.
Internal firmware can be updated by user. It is recommended check for firmware updates every 6 months
or so to get the latest UI improvements, new features and bug fixes. To do this:
1.
Go to
www.meatest.com/drivers-updates
and find firmware update file (.upl) for your device.
If your device is not listed, contact Meatest support and ask for firmware update file for your
device.
2.
Check if the update file version is greater than yours, installed FW version can be found in
MENU > Information > Software version. If the installed version is the same, stop the update
process.
3.
Download Uploader program from the same website and install it if you haven’t already.
4.
Connect the instrument to your PC using RS232 cable and use Uploader to update your
instrument's firmware. The calibrator will be restarted once the update is finished.
5.
Check installed FW version.
Summary of Contents for 9010+
Page 1: ......
Page 2: ......
Page 3: ......
Page 4: ......
Page 5: ......
Page 33: ...Following items can be modified ...
Page 46: ......
Page 47: ......
Page 48: ......
Page 49: ......
Page 50: ......
Page 51: ......
Page 55: ......
Page 57: ......
Page 58: ......
Page 78: ...Temperature range 250 00 1820 00 C Cold junction comp modes Manual ...