Measuretek PS-103-180 Operation Manual Download Page 4

 
 

PS

103

 

www.measuretek.net

 

.

 

2

 

.

 

 
 

Display 

 

>0< - Scale is zeroed and gross weight is zero. 

 

Net - Display reading is net weight; tare is not 0.   

 

lb, kg, oz - Unit of measure. 

 

Hold - Scale is in dynamic weighing mode.   

-

 

Hold flashes - actual fluctuating weight displayed. 

-

 

Hold does not flash - locked weight is displayed.   

 

 
 
Function Keys 

KEY 

MODE 

DEFINITION 

PRINT/ HOLD 

Weighing mode

<3 seconds 

Sends output data via the USB port 

>3 seconds 

Enters or exits HOLD mode 

Setup or Calibration mode 

Shifts the flashing data entry position from right to left 

UNIT  

Weighing mode 

Chooses weight unit of measure 

Setup/Calibration mode 

Increases the digit in the flashing data entry position by 1 

TARE 

Weighing mode 

Tares the weight 

Setup or Calibration mode 

Confirms the input data and forwards to next step   

ON/OFF/ 

ZERO 

Weighing mode

<3 seconds 

Zeros the platform weight   

>3 seconds 

Powers off the scale 

Setup or Calibration mode 

Exits to normal weighing mode 

ON/OFF/ 

ZERO 

UNIT 

Weighing mode 

(more than 3 seconds) 

Enter user parameter setup mode. 

ON/OFF/ 

ZERO 

TARE 

Weighing mode 

(more than 3 seconds) 

Enter calibration mode 

 
 

Contents   

 

Indicator 

 

Scale platform 

 

Cable from load cell to indicator 

 

6’ USB cable 

 

AC120V/DC6V 500mA UL adaptor 

 

Owner’s Manual 

Summary of Contents for PS-103-180

Page 1: ...W W W M E A S U R E T E K N E T PS 103 Operation Manual Contents subject to change without notice Version 1 0 11 2011 ...

Page 2: ......

Page 3: ...forms poorly try moving to a different room or location Avoid using in heavy vibration or heavy airflow conditions Specifications Model PS 103 35 PS 103 68 PS 103 75 PS 103 180 PS 103 200 Max Capacity 77lb 35kg 150lb 68kg 165lb 75kg 397lb 180kg 440lb 200kg Readability 0 05lb 0 02kg 0 5lb 0 2kg 0 1lb 0 05kg 0 5lb 0 2kg 0 2lb 0 1kg Resolution 1 1750 1 300 1 1500 1 900 1 2000 Construction Epoxy paint...

Page 4: ...from right to left UNIT Weighing mode Chooses weight unit of measure Setup Calibration mode Increases the digit in the flashing data entry position by 1 TARE Weighing mode Tares the weight Setup or Calibration mode Confirms the input data and forwards to next step ON OFF ZERO Weighing mode 3 seconds Zeros the platform weight 3 seconds Powers off the scale Setup or Calibration mode Exits to normal ...

Page 5: ... The default calibration input weight will be 50 of the scale s rated capacity Use the PRINT HOLD and UNIT keys adjust to the standard calibration weight you will use Note do not use less than 10 of the scale s rated capacity Place the standard calibration weight on the platform and press TARE to confirm your input and move to the next step If CAL Er is displayed press ZERO ON OFF key to exit this...

Page 6: ...ight within 3 seconds Note When weight exceeds 10d and the variation is within 2d 50d the stable weight is displayed until less than 10d is detected on the platform 1d 1 division 1 unit of readability as noted in Specifications U5 1 Serial communication output format 0 Communication is disabled 1 Outputs stable weight unit and status when PRINT HOLD is pressed but does not receive data 2 Output gr...

Page 7: ...et reading minus decimal point unit CR LF H1H2H3 CR ETX Bytes used 8 Weight reading 1 Minus 1 Decimal point 2 or 5 Weight unit 3 Current status H1 H2 H3 U5 3 Continuously outputs the current displayed reading weight unit and current status data but will not receive data The output format is the same as U5 1 U5 4 Continuously outputs the current gross weight tare weight net weight data weight unit ...

Page 8: ...e character indicates positive weight The minus sign follows after the first digit Leading zero digits are suppressed Serial Communication Symbols LF Line Feed character hex 0AH CR Carriage Return character hex 0DH ETX End of Text character hex 03 SP Space hex 20H H1H2H3 Three status bytes see below for byte definitions p Polarity minus sign for negative weight and space for positive weight W1 W8 ...

Page 9: ...F H1H2H3 CR ETX Zero function is activated and returns to current scale status like pressing ZERO ON OFF If zero function cannot be activated it will return to current scale status Command T CR 54h 0dh Response LF H1H2H3 CR ETX Tare function is activated and returns to current scale status like pressing TARE If tare function cannot be activated it will return to current scale status Command U CR 5...

Page 10: ... 3 6V Replacement Parts Part Number Description MH12R97601G AC120V DC6V 500mA UL adaptor One Year Limited Warranty MeasureTek products covered in this manual are guaranteed to be free from defects in material and workmanship for a period of one year after date of purchase Misuse accidental damage overload alteration and improper installation are expressly excluded Any product which is determined t...