background image

USB-3103 User's Guide 

Functional Details 

17 

Caution!

 

The +5V terminal is an output. Do not connect to an external power supply or you may damage 
the USB-3103 and possibly the computer. 

Synchronizing multiple units 

You can connect the 

SYNCLD

 terminal pin (pin 49) of two USB-3103 units together in a master/slave 

configuration and simultaneously update the DAC outputs of both devices. Do the following. 

1.

Connect the SYNCLD pin of the master USB-3103 to the SYNCLD pin of the slave USB-3103.

2.

Configure the SYNCLD pin on the slave device for input to receive the D/A LOAD signal from the master
device. Use 

Insta

Cal to set the direction of the SYNCLD pin.

3.

Configure the SYNCLD pin on the master device for output to generate an output pulse on the SYNCLD
pin.

Set the Universal Library 

SIMULTANEOUS

 option for each device. Refer to the 

Universal Library User's Guide

 

for information on how to configure the USB-3103 with the Universal Library. 

When the SYNCLD pin on the slave device receives the signal, the analog output channels on each device are 
updated simultaneously. 

An example of a master/slave configuration is shown here. 

S

Slave

PMD-1608F

Master

device

Slave

device

SYNCLD pin

(terminal pin 49)

D/A LOAD

 signal

Configure the SYNCLD 

pin for output

Configure the SYNCLD 

pin for input

Figure 8. Synchronizing the update of multiple devices 

Summary of Contents for USB-3103

Page 1: ...USB 3103 Eight analog voltage output 16 bit Eight digital 1 0 one 32 bit event counter Full speed USB 2 0 device User s Guide F A MEASUREMENT y COMPUTING...

Page 2: ...USB 3103 USB based Analog Output User Guide Document Revision 3A March 2009 Copyright 2009 Measurement Computing Corporation...

Page 3: ...in any form by any means electronic mechanical by photocopying recording or otherwise without the prior written permission of Measurement Computing Corporation Notice Measurement Computing Corporation...

Page 4: ...No license is granted by implication or otherwise under any patent or copyrights of Measurement Computing Corporation All rights reserved No part of this publication may be reproduced stored in a retr...

Page 5: ...103 to your system 11 Calibrating the USB 3103 12 Chapter 3 Functional Details 13 External components 13 USB connector 13 Status LED 13 Power LED 13 Screw terminal banks 14 Analog voltage output termi...

Page 6: ...USB 3103 User s Guide 6 Mechanical 22 Main connector and pin out 22 Declaration of Conformity 23...

Page 7: ...of objects on a screen such as buttons text boxes and check boxes italic text Italic text is used for the names of manuals and help topic titles and to emphasize a word or phrase Where to find more in...

Page 8: ...DAC You set the voltage output range of each DAC channel independently with software for either bipolar or unipolar The bipolar range is 10 V and the unipolar range is 0 to 10 V The analog outputs may...

Page 9: ...annel 32 bit 1 DIO 8 channels 8 USB Full speed USB 2 0 Compliant Interface SPI D A 8 4 D A 4 SYNCLD 1 4 channel Quad Analog Voltage Output 8 channels 16 bit Figure 2 USB 3103 block diagram Software fe...

Page 10: ...Start pdf This booklet supplies a brief description of the software you received with your USB 3103 and information regarding installation of that software Please read this booklet completely before...

Page 11: ...ows Update or download the update from www microsoft com downloads details aspx familyid 733dd867 56a0 4956 b7fe e85b688b7f86 displaylang en For more information refer to the Microsoft Knowledge Base...

Page 12: ...ion disconnect the USB cable from the computer and then reconnect it This should restore communication and the power LED should turn back on Calibrating the USB 3103 The USB 3103 is shipped fully cali...

Page 13: ...ower and communication to the USB 3103 The voltage supplied through the USB connector is system dependent and may be less than 5 V No external power supply is required Status LED The Status LED indica...

Page 14: ...03 pins 1 to 28 provide the following connections Four analog voltage output connections VOUT0 VOUT2 VOUT4 VOUT6 Four analog ground connections AGND Eight digital I O connections DIO0 to DIO7 Screw te...

Page 15: ...software programmable for either bipolar or unipolar The bipolar range is 10 V and the unipolar range is 0 to 10 V The channel outputs may be updated individually or simultaneously Digital I O termina...

Page 16: ...as an input slave mode to receive the D A LOAD signal from an external source When the SYNCLD pin receives the trigger signal the analog outputs are updated simultaneously SYNCLD pin must be logic lo...

Page 17: ...gnal from the master device Use InstaCal to set the direction of the SYNCLD pin 3 Configure the SYNCLD pin on the master device for output to generate an output pulse on the SYNCLD pin Set the Univers...

Page 18: ...50 mS typ Amplitude 5 V peak typ Differential non linearity Note 3 Calibrated 1 25 LSB typ 2 LSB to 1 LSB max Un calibrated 0 25 LSB typ 1 LSB max Output current VOUTx pins 3 5 mA typ Output short cir...

Page 19: ...cy specifications Range Relative accuracy LSB 10 V 0 to 10 V 4 0 typ 12 0 max Analog output calibration Table 5 Analog output calibration specifications Parameter Conditions Specifications Recommended...

Page 20: ...V min No load 3 8 V min Output low voltage Note 6 IOL 2 5 mA 1 1 V max No load 0 6 V max Note 5 SYNCLD is a Schmitt trigger input and is over current protected with a 200 Ohm series resistor Note 6 Wh...

Page 21: ...utput current Note 9 Available at terminal block pin 56 10 mA max Note 7 This is the total quiescent current requirement for the USB 3103 which includes up to 10 mA for the status LED This does not in...

Page 22: ...ain connector specifications Connector type Screw terminal Wire gauge range 16 AWG to 30 AWG Pin Signal Name Pin Signal Name 1 VOUT0 29 VOUT1 2 NC 30 NC 3 VOUT2 31 VOUT3 4 NC 32 NC 5 AGND 33 AGND 6 VO...

Page 23: ...the following conditions must be met The host computer peripheral equipment power sources and expansion hardware must be CE compliant Equipment must be operated in a controlled electromagnetic enviro...

Page 24: ...ion 10 Commerce Way Norton Massachusetts 02766 508 946 5100 Fax 508 946 9500 E mail info mccdaq com www mccdaq com NI Hungary Kft H 4031 Debrecen H tar t 1 A Hungary Phone 36 52 515400 Fax 36 52 51541...

Reviews: