Table 15. RMS noise performance specifications (µVRMS)
A/D data rate
Range (V)
3750 S/s 2000 S/s 1000 S/s 500 S/s 100 S/s 60 S/s 50 S/s 25 S/s 10 S/s 5 S/s 2.5 S/s
±20
34.90
28.32
19.22
15.17
8.67
8.67
7.51
7.22
3.61
3.61 2.17
±10
19.22
15.17
10.87
6.83
4.62
4.62
4.05
2.89
1.81
1.44 1.44
±5
8.60
7.21
5.18
3.92
2.46
2.17
2.17
2.16
0.90
0.72 0.72
±2.5
4.99
4.36
2.72
2.15
1.14
1.07
1.07
0.87
0.58
0.43 0.36
±1.25
2.81
2.66
2.10
1.41
0.83
0.83
0.76
0.58
0.51
0.51 0.43
±0.625
2.86
2.51
1.28
1.12
0.81
0.73
0.66
0.58
0.38
0.40 0.30
±0.3125
2.32
2.24
1.24
1.05
0.72
0.71
0.68
0.56
0.51
0.39 0.37
±0.15625
2.01
2.55
1.13
1.00
0.86
0.68
0.68
0.64
0.40
0.47 0.35
±0.078125 2.04
2.28
1.39
1.04
0.65
0.63
0.60
0.54
0.35
0.32 0.36
Table 16. Noise-free resolution specifications (bits)
A/D data rate
Range (V) 3750 S/s 2000 S/s 1000 S/s 500 S/s 100 S/s 60 S/s 50 S/s 25 S/s 10 S/s 5 S/s 2.5 S/s
±20
17.4
17.7
18.2
18.6
19.4
19.4
19.6
19.6
20.6
20.6 21.4
±10
17.2
17.6
18.1
18.7
19.3
19.3
19.5
20.0
20.6
21.0 21.0
±5
17.4
17.6
18.1
18.5
19.2
19.4
19.4
19.4
20.6
21.0 21.0
±2.5
17.2
17.4
18.1
18.4
19.3
19.4
19.4
19.7
20.3
20.7 21.0
±1.25
17.0
17.1
17.4
18.0
18.8
18.8
18.9
19.3
19.5
19.5 19.7
±0.625
16.0
16.2
17.1
17.3
17.8
18.0
18.1
18.3
18.9
18.8 19.2
±0.3125
15.3
15.3
16.2
16.4
17.0
17.0
17.0
17.3
17.5
17.9 17.9
±0.15625
14.5
14.1
15.3
15.5
15.7
16.1
16.1
16.1
16.8
16.9 17.1
±0.078125 14.5
14.3
15.0
15.4
16.1
16.2
16.2
16.4
17.0
17.1 16.9
Channel switching error
Table 17. Step response accuracy specifications
Accuracy S/s
Range (V)
3750
S/s
2000
S/s
1000
S/s
500
S/s
100
S/s
60 S/s
50 S/s
25 S/s
10 S/s
5 S/s
2.5 S/s
±20
0.001
1%
0.0006
%
0.0004
%
0.0003
%
0.0002
%
0.0002
%
0.0002
%
0.0002
%
0.0001
%
0.0001
%
0.0001
%
±10
0.001
0%
0.0008
%
0.0005
%
0.0004
%
0.0002
%
0.0002
%
0.0003
%
0.0002
%
0.0001
%
0.0001
%
0.0001
%
±5
0.000
9%
0.0008
%
0.0004
%
0.0004
%
0.0003
%
0.0002
%
0.0002
%
0.0002
%
0.0001
%
0.0001
%
0.0001
%
±2.5
0.001
0%
0.0007
%
0.0008
%
0.0004
%
0.0003
%
0.0002
%
0.0002
%
0.0002
%
0.0002
%
0.0001
%
0.0001
%
±1.25
0.001
3%
0.0009
%
0.0008
%
0.0007
%
0.0004
%
0.0004
%
0.0003
%
0.0003
%
0.0003
%
0.0003
%
0.0003
%
±0.625
0.002
2%
0.0016
%
0.0011
%
0.0011
%
0.0007
%
0.0007
%
0.0005
%
0.0005
%
0.0004
%
0.0005
%
0.0003
%
±0.3125
0.003
1%
0.0031
%
0.0020
%
0.0017
%
0.0015
%
0.0012
%
0.0010
%
0.0010
%
0.0012
%
0.0009
%
0.0009
%
±0.15625
0.005
6%
0.0062
%
0.0048
%
0.0037
%
0.0032
%
0.0025
%
0.0024
%
0.0021
%
0.0019
%
0.0022
%
0.0016
%
±0.078125
0.011
4%
0.0123
%
0.0076
%
0.0070
%
0.0041
%
0.0051
%
0.0046
%
0.0036
%
0.0032
%
0.0030
%
0.0034
%
Channel switching error is defined as the accuracy that can be expected after one conversion when switching
from a channel with a DC input at one extreme of full scale to another channel with a DC input at the other
extreme of full scale, expressed in terms of percentage of full scale value.