PCI-DAS6052 User's Guide
Specifications
Table 6-2. Absolute Accuracy Components – All values are (±)
Noise +Quantization
(µV)
Range
% of
Reading
Offset
(µV)
Single Pt Averaged
1
Temp Drift
(%/DegC)
Absolute
Accuracy at
FS (mV)
±10V
0.0371 947 981
87.0
0.0006 4.747
±5V
0.0071 476 491
43.5
0.0001 0.876
±2.5V 0.0371 241 245
21.7
0.0006 1.190
±1V 0.0371
99.2 98.1 8.7
0.0006 0.479
±500mV
0.0371 52.1 56.2
5.0
0.0006 0.243
±250mV 0.0421
28.6
32.8
3.0
0.0006
0.137
±100mV 0.0471
14.4
22.4
2.1
0.0006
0.064
±50mV
0.0471 9.7 19.9
1.9
0.0006 0.035
0 to 10V
0.0071 476 491
43.5
0.0001 1.232
0 to 5V
0.0371 241 245
21.7
0.0006 2.119
0 to 2V
0.0371
99.2
98.1
8.7
0.0006
0.850
0 to 1V
0.0371
52.1
56.2
5.0
0.0006
0.428
0 to
500mV
0.0421 28.6 39.8
3.0
0.0006 0.242
0 to
200mV
0.0471 14.4 22.4
2.1
0.0006 0.111
0 to
100mV
0.0471 9.7 19.9
1.9
0.0006 0.059
1. Averaged measurements assume averaging of 100 single-channel readings
6-3