Chapter 8
168
14.
In the Range box, select
PGH
Zero
.
15.
If the displayed value is not 0.0000 V (within 0.001 V), perform the procedure in the next
section; otherwise, click
Quit
when you are finished calibrating the analog input circuitry.
Once you have finished this procedure, continue with
“Calibrating the Analog Output
Note:
If you are not satisfied with the analog input calibration, you can load the factory
default settings stored in the EEPROM by clicking
Restore
in the Factory Settings box.
Calibrating the PGH Zero Setting
PGH Zero is a factory-calibrated setting and, generally, should not need adjustment. However,
if you select
PGH Zero
in the Range box and a value other than 0.0000 V is displayed, do the
following to calibrate this setting:
1.
In the Range box, select
PGH
Zero
.
2.
Physically adjust potentiometer R2 (labelled PGZ) on the DT3016 board until the display
reads 0.0000 V (0.001 V).
shows the location of this potentiometer.
Figure 77: Location of Potentiometer R2 on the DT3016 Board
3.
Click
Quit
when you are finished.
Once you have finished this procedure, continue with the next section.
Potentiometer for PGH Zero (PGZ)
R2
Summary of Contents for Data Translation DT3010 Series
Page 2: ...DT3010 Series UM 16866 V User s Manual Title Page ...
Page 5: ......
Page 15: ...About this Manual 14 ...
Page 16: ...15 1 Overview Features 16 Supported Software 18 Accessories 19 Getting Started Procedure 21 ...
Page 23: ...Chapter 1 22 ...
Page 24: ...Part 1 Getting Started ...
Page 25: ......
Page 37: ...Chapter 2 36 ...
Page 49: ...Chapter 3 48 ...
Page 91: ...Chapter 4 90 ...
Page 102: ...Part 2 Using Your Board ...
Page 103: ......
Page 147: ...Chapter 6 146 ...
Page 159: ...Chapter 7 158 ...
Page 181: ...Chapter 9 180 ...
Page 193: ...Appendix A 192 ...
Page 213: ...Appendix B 212 ...
Page 231: ...Index 230 ...