Page 16
Installation and Assembly
ACC5703
Selection of measurement axis (x, y, z) at every channel possible
Off (zero output)
Continuous mode, current 4 - 20 mA
Continuous mode, voltage 0.5 - 4.5 V
Switching mode, voltage 0 - 5 V
Fig. 7 Operation modes of the analog output channels
4.4.1
Continuous Operation Mode
The sensor makes the acceleration value available as analog output variable either as current or voltage
value on separate pins, depending on the configuration using the software tool of Micro-Epsilon.
In this process, the symmetrical measurement range in the unit g is scaled to the respective analog range.
The sensitivity increases with decreasing measurement range as only a small acceleration range is scaled to
the same output range
8
12
4
6
16
20
10
14
18
Current
in mA
Acceleration in g
-2 g
-4 g
-8 g
0 g
+2 g
+4 g
+8 g
0
0.5
1
2
3
4
5
1.5
2.5
3.5
4.5
Voltage in V
Acceleration in g
-2 g
-4 g
-8 g
0 g
+2 g
+4 g
+8 g
Fig. 8 Scaling of the acceleration measurement range to analog output variable current or voltage