OPERATING PROCEDURE 

 
To bring the table to the starting position, pull out the Table Release Knob, slide it to the left and release it.  
Place the specimen on the centre of the table, switch on and press the Reset Switch.  The speed may be adjusted 
with the Speed Control Knob – clockwise to increase.  After the tissue has been cut into slices, the machine will 
automatically stop when the table reaches the end of its travel.  If it is required to cut the tissue into cubes, the 
table may be rotated through 90º and the procedure repeated. 
 
If the blade holder is removed for sterilising, we recommend sliding a piece of paper under the blade holder to 
prevent the nuts and washers dropping into the instrument. 
 
After sterilising the cutting table, apply a little paraffin wax to the underside before replacing it on the 
mounting disc. 
 

 

 

SPECIFICATION 

 

 

RATING 

 220-240 volts 50 hz        0.5 amp 

 

 

 or 100-115 volts 60 hz    1 amp 

 
 
FUSES 

 3.15 amp. Anti-surge (T) 5 x 2Omm  

 
 

 

 

 
OPERATING CONDITIONS 

 Indoor use 

 

 

 Altitude up to 2000 metres 

 

 

 Ambient temperature 5º C

 

to 40º C 

 

 

 Maximum RH 80% at 31º C decreasing linearly 

 

 

 to 50% RH at 40º C 

 

 

 Installation Category II 

 

 

 Pollution degree I or 2 (IEC 664) 

 
 

DIMENSIONS 

 31 x 28.5 x 15 cms 

 

WEIGHT : 

 

6.3 

kg 

 

 

 

 

 
 
 