OPERATING PROCEDURE
To bring the table to the starting position, pull out the Table Release Knob, slide it to the left and release it.
Place the specimen on the centre of the table, switch on and press the Reset Switch. The speed may be adjusted
with the Speed Control Knob – clockwise to increase. After the tissue has been cut into slices, the machine will
automatically stop when the table reaches the end of its travel. If it is required to cut the tissue into cubes, the
table may be rotated through 90º and the procedure repeated.
If the blade holder is removed for sterilising, we recommend sliding a piece of paper under the blade holder to
prevent the nuts and washers dropping into the instrument.
After sterilising the cutting table, apply a little paraffin wax to the underside before replacing it on the
mounting disc.
SPECIFICATION
RATING
:
220-240 volts 50 hz 0.5 amp
or 100-115 volts 60 hz 1 amp
FUSES
:
3.15 amp. Anti-surge (T) 5 x 2Omm
OPERATING CONDITIONS
:
Indoor use
Altitude up to 2000 metres
Ambient temperature 5º C
to 40º C
Maximum RH 80% at 31º C decreasing linearly
to 50% RH at 40º C
Installation Category II
Pollution degree I or 2 (IEC 664)
DIMENSIONS
:
31 x 28.5 x 15 cms
WEIGHT :
6.3
kg