a.
Turn the base unit on and set the voltage to 3.0 to 3.5kV
b.
Slowly push the Bi-Polar forceps away from you. Go from the tip of the forceps to the base.
c.
Repeat using the second tine.
d.
Turn the Bi-Polar forceps over and repeat the test of both tines
e.
Alarm will sound and LED will Flash if a fault is found in the coating which will indicate a fault with the instrument
1.
Insert wire brush or ring or Tri-hole electrode firmly into the Red port on the top of the base unit (red port) (optional)
1.
Attach the clamp end of the (Green) ground wire to conductive core of instrument under examination. See Picture A, B and C for illustrations.
Using the Optional HV RED Lead (PN MMRWP-006 not included in kit)
1.
Insert the Red HV lead probe wire firmly into the red port - top of the base unit.
2.
Insert wire saddle, brush, ring or Tri-hole electrode firmly into end of yellow handle
and
repeat steps 9-12
The Lithium Polymer Battery can only be replaced at the McGan facility.
DO NOT attempt to replace the Battery
Use the brush to check the handle
of the electrosurgical instrument
Picture B
Picture A
Exposed end
Attach ground clamp here
OPTIONAL Assembly