6.1 Visual Inspection
Visually inspect the unit and cable assemblies for the following:
Device, housing and mains cable (not damaged or cracked)
Electrode cable sheathing and connectors (undamaged)
No cracks, abrasion or wear in any cable assembly
Defective units or damaged cables must be replaced immediately.
Warning
5.2 Starting a Recording
1 Preparing the patient
2 Apply electrodes
3 Ask the patient about their well-being
4 Performing a measurement
5 In case suction pulbs and extremity electrodes are applied: Clean them before next use
6 Maintenance and Care
The device requires regular checks (see page 21, Chapter 6.4). The test results must be recorded in
writing and compared to the values in the accompanying documents.
Maintenance work not described in this section may only be performed by a qualified,
authorised technician.
The following table indicates the intervals and responsibilities of the maintenance work required.
Local regulations in your country may stipulate additional or different inspection intervals and tests.
Interval
Maintenance step
Responsible
Before every use
Visual inspection of the device and ECG electrodes
User
Every 6 months
Visual inspection of the device
(see page 21, 6.4 Inspection report)
- Cables and accessories
Functional tests according to the instructions
(see page 21, 6.4 Inspection report)
User
Every 12 months
Safety test according to § 11 MPBetreibV
qualified staff