Sterilization:
This device is not sterile,
Accessories are not sterile
9
Malfunction
If you notice a malfunction that is not commented on in the documents accompanying the
device, please inform your distributor or the manufacturer.
In the case of a shipment of the device, please observe the following instructions:
▪
Decontaminate and clean the unit and its accessories.
▪
Use the original packaging, including the retaining flanges.
▪
Attach all accessories to the unit.
▪
Set up the different elements.
▪
Make sure the package is properly closed.
Shipping address:
Electronique du Mazet
ZA Route de Tence
43520 Le Mazet St Voy
Tel : (33) 4 71 65 02 16
Fax: (33) 4 71 65 06 55
Email: [email protected]
Possible operating anomalies:
Description of the anomaly
Possible causes
Actions
No connection to Biostim
Bluetooth connection problem
-Check that the Biomoov is charged
-Switching the Biomoov off and on
-Click on the Bluetooth logo (under the house).
No charging of the device
USB cord problem
Make sure the cord is not damaged
Flat trace in biofeedback
windows
- loss of communication with the
module.
- no sensor on the considered
input
- exit the current treatment and return to the
main office.
- check the path used
If the device is dropped or if water penetrates, it is imperative to have the device checked by
Électronique du Mazet to exclude any risk (patient and user) related to the use of the device.