Installation
- 16 -
Figure 12 Align and secure the equipment
3.2 Electrical connections
DANGER
Only qualified experts should conduct the tasks described in this section.
DANGER
Do not connect the electrical supply to the mains if the equipment has not
been wired and fused correctly.
Sufficiently protect the electrical power supply against short circuits.
For the external power supply, always connect a residual-current circuit
breaker·(trip current max.: 30 mA) between the mains and the system.
If the equipment is to be installed outdoors, place an overload protection
between mains and system.
If the mains plug·of the power supply cable·is removed, a suitable double-
pole one-way switch must be installed immediately next to the display
unit·with clear labeling for the power supply.
Products intended by the manufacturer for outdoor use offer a higher
level of protection against the penetration of liquids and dust. If these
products are connected to a mains socket with a cable and
plug·rather than a permanently connected cable, the plug and socket
are much more susceptible to liquid and dust penetration. The
operator must sufficiently protect the plug and outlet against liquid
and dust penetration in accordance with local safety regulations. If the
instrument is to be used outdoors, it must be connected to a suitable
outlet with a protection type of at least IP44 (splash protection).
Summary of Contents for SP5 A
Page 12: ...Installation 12 3 1 2 Select place of installation Figure 5 Select place of installation...
Page 22: ...Installation 22 Figure 23 Screw in the union nut Figure 24 Close the lid...
Page 35: ...Operation 35 Figure 45 Remove full bottles Figure 46 Replace with empty bottles...
Page 50: ...Maintenance and cleaning 50 5 6 7...
Page 55: ...Chapter 6 Spare parts and accessories 55 6 1 Spare parts Figure 74 Plastic dosing vessel...
Page 56: ...Spare parts 56 Figure 75 Glass dosing vessel 350 ml...
Page 57: ...Ersatzteile 57 Figure 76 Glass dosing vessel 500 ml...
Page 58: ...Spare parts 58 Figure 77 Glass dosing vessel flow through...