PRINCIPLE OF OPERATION:
The UltraMaxO
2
Oxygen Analyzer measures oxygen concentration and flow using
ultrasound technology and measures pressure using a piezoresistive silicon
pressure sensor.
WARNINGS:
Indicates a potentially hazardous situation, if not avoided, could result in death or
serious injury.
•
Not for use in an MRI environment.
•
Improper use of the UltraMaxO
2
can cause inaccurate oxygen readings leading to
improper treatment and/or patient harm. Follow the procedures outlined in this
user manual.
•
The UltraMaxO
2
is for checking oxygen concentrators only.
•
DO NOT use the UltraMaxO
2
for continuous oxygen monitoring.
•
DO NOT use the UltraMaxO
2
to measure the oxygen concentration of a concentrator
when flowing at rates lower than its optimal performance as specified by the
concentrator manufacturer; generally 4 LPM or less on concentrators that have a
maximum flow of 10 LPM, and 1 LPM or less on concentrators that have a maximum
flow of 5 LPM.
•
Not for use in anesthesia applications or for measuring oxygen concentration from
any sources other than conventional oxygen concentrators.
•
Not for use with inhalation agents. Operating the UltraMaxO
2
in flammable or
explosive environments may result in fire or explosion.
•
Not suitable for use in the presence of flammable anesthetic mixtures.
•
Users must become thoroughly familiar with the information contained in this
Operation Manual before use. Strict adherence to the operating instructions is
necessary for safe, effective product performance. This product will perform only as
designed if operated in accordance with the manufacturer’s operating instructions.
•
Use only genuine Maxtec accessories. Failure to do so may seriously impair the
performance of the UltraMaxO
2
. Repair or alteration of the UltraMaxO
2
by anyone
other than an authorized Maxtec service representative could cause the product to
fail to perform as designed.
•
Use of the UltraMaxO
2
near devices that generate electrical fields may cause
erratic readings.
•
If the UltraMaxO
2
is ever exposed to liquids from spills or immersion, immediately
remove the batteries and let the device dry completely. When dry, replace the
batteries and check for proper operation.
•
DO NOT
autoclave or expose the UltraMaxO
2
to high temperatures (>60°C).
•
DO NOT
use ethylene oxide sterilization. performance in flow and pressure readings.
!