![Max-Ability Pressalit Aquamax User Manual Download Page 23](http://html1.mh-extra.com/html/max-ability/pressalit-aquamax/pressalit-aquamax_user-manual_1743049023.webp)
Electromagnetic compatibility
Electromagnetic immunity
Test standard
Compliance
level
Test level /
details
Test
results
EN60601-1-2
and EN17966
(Home healthcare
environment)
=> 10V/m
2,7GHz
15V/m
Frequency range:
80MHz- 6GHz
1% step size
Passed
EN60601-1-2
(Proximity fields
from RF wireless
communications
equipment)
=> 10V/m
2,7GHz
15V/m
Frequency range:
385 - 5785 MHz
Passed
IEC 61000-4-4
(Electrical fast
transients /
bursts)
Quick pulse simulation Passed
Conducted
disturbances
induced by RF
Pulse
subjection, 6V
Pulse
subjection, 10V
Passed
Electrostatic
discharge
Statically charged
person discharge on
unit, 8kV and 15kV
Passed (*)
Test standard
Test level / details
Test results
60601-1-2 ch . 7 .1 .7 -
CISPR 14
30MHz – 1GHz
Passed
Electromagnetic emissions
* If a statically charged person discharges on or near the shower trolley control
box, a fault may occur in the lifting columns, and a reset must be performed . See
"
Resetting
" .
max-ability.com 800.577.1555 [email protected]