10
About the operation with the OUTPUT switch on the front panel or a remote switch, and the indication of
OUTPUT LED
The procedure of operation for an output is different with whether the remote switch is set to ON or OFF at
the time of POWER ON.
About the difference arising from the state of the remote switch at the time of POWER ON
When a blackout protection is ON
When the POWER switch is turned off during output preparation or an output, a blackout protection operates.
When the POWER switch is turned on again, it will be in an output halt condition.
① In the case of the remote switch OFF and the POWER switch ON
② In the case of the remote switch ON and the POWER switch ON
*EPR series is high voltage power supply. Therefore, for the safety of users, at the time of factory shipments,
it is a setup switch is not outputted unless the OUTPUT switch is turned on. When a blackout protection is
OFF, output can be done only with the operation of the remote switch even if the OUTPUT switch is not
turned on.
OUTPUT LED
Lights-out
OUTPUT LED Blinking
OUTPUT LED
Lights-out
OUTPUT LED Lighting
OUTPUT LED Lighting
OUTPUT LED
Lighting:in an output condition
Blinking:in an output preparatory
condition
Lights-out : in an output halt
condition
POWER ON
OUTPUT ON
Remote switch OFF
OUTPUT ON
OUTPUT LED Lighting
OUTPUT LED
Lights-out
OUTPUT LED
Lights-out
Remote switch ON
OUTPUT LED Blinking
OUTPUT LED
Lights-out
Remote switch ON
OUTPUT LED Lighting
POWER ON
OUTPUT ON
Remote switch ON
OUTPUT ON
Remote switch ON
3