7
COMM.PARITY
Set communication verification method
NONE
No calibration
EVEN
Even parity
ODD
Odd parity
ADDRESS SET
Set the load communication address
1~200
KEY LOCK SET
Set the keyboard to unlock the password (null or 0
means no password)
EXIT
SYSTEM SET
MAX CURRENT SET
Set the software current limit, if the current limit is
greater than 1/10 of the maximum current, the current
is high range, otherwise low range (such as 120A
machine, 12A following low range)
MAX VOLTAGE SET
Set the voltage limit, if the voltage limit is greater than
20V, the voltage is high range, or low range
MAX POWER SET
Set the power cap
TERMINAL SEL
Input terminal selection
FRONT
Select front panel terminal input
BACK
Select the rear panel terminal input
EXIT
LIST
LOAD LIST
Select LIST file, 1 ~ 8
EDIT LIST
Edit LIST file
MINIMUM TIME
Editing the minimum time fineness (0.02 ~ 1310.7mS)
LIST MODE
LIST output mode
CONTINOUS
Continuous output
END HOLD
The end of the output holds the last output level
END RESET
End of output to maintain no-load output
STEP LENGTH
Total steps (1 ~ 200)
STEP n
1 ~ total step
CURRENT
Set the current
TIME
duration
EXIT
AUTO
TEST
LOAD AUTO TEST
Select automatic test file 1 ~ 8
EDIT AUTO TEST
Edit automatic test files
STEP LENGTH
Set the total step size
STEP n
WORK MODE
LOAD OFF MODE
No-load mode
CC MODE
Constant current mode
CV MODE
Constant voltage mode
CP MODE
Constant power mode
CR MODE
Constant resistance mode
SHORT MODE
Short circuit mode
TEST MODE
Conformity test mode