Rollex Medical
–
MATOS ARIA Eco 150L User Manual v.02 24/04/2018
Product Information
Important Information
The refrigerator contains environment-friendly, non-ozone depleting refrigerant R600a. As R600a is a
flammable gas, it is important to avoid damage to the refrigeration circuit during transport and
installation. If the refrigeration circuit is damaged, avoid using a naked flame in the vicinity of the
refrigerator and connecting power to the refrigerator. Also make sure there is good ventilation in the
room. If you are in doubt please contact your supplier.
WARNING:
Keep ventilation openings, in the appliance enclosure or in the built-in structure, clear of
obstruction.
WARNING:
Do not use mechanical devices or other means to accelerate the defrosting process, other
than those recommended by the manufacturer.
WARNING:
Do not damage the refrigerant circuit
WARNING:
Do not use electrical appliances inside the food storage compartments of the appliance,
unless they are of the type recommended by the manufacturer.
Product Specifications
This refrigerator has been designed and built with the intention of the product being used within a
Pharmacy and is designed to store contents b2°C to +8°C.