Master equipment TP15140 Instructions Download Page 2

CARE & MAINTENANCE:

Unplug table. Wipe off tabletop with a sponge or damp cloth and mild cleaner after use. Use mild soap and water, or a 
non-caustic cleaner. Tabletop can also be disinfected using a diluted solution of Top Performance® Cherry-256™, Lemon-256™, 
Wintergreen-256™, or a diluted bleach and water solution. Periodically vacuum or dry wipe lifting mechanism to keep it free of 
hair and debris. Base can also be cleaned when needed with damp cloth or sponge. Dry off any excess water. Check electric cord 
and plug for signs of damage or wear.  

TROUBLESHOOTING:

In rare cases, some electrical connections may be jostled loose during
shipment. If controls do not operate once the table is plugged in,
please read the following steps for either checking and reconnecting a
loose wire in the control box or replacing the transmission shaft.

Tools Required: Phillips Screwdriver, Open End Wrench, Allen Wrench

Problem: Table Does Not Turn On.

Possible Causes: Table Not Plugged In, Control Box
Connections Loose, Transformer Defective.

1. Check that table is plugged into a working outlet.
2. Check connections in Control Box (Diagram 1, A).

To Check Control Box (A):

1. Unplug table and turn it on its side so that black wire cover panel on side of frame is facing top and easily accessible. Using a

phillips screwdriver, unscrew the six screws along the side of the bottom panel of the Control Box (A) to uncover transformer. 

2. Make sure both of the “M” Plugs (A-1) (black wires) are securely fastened 

to the transformer socket.

3. Ensure all plug connections for the foot pedals (gray wires) are 

securely fastened  (A-3).

4. Check Transformer for damage. If damaged, unscrew two screws on the 

top panel of the Control Box and remove metal frame from around
Transformer (A-2).

5. To replace Transformer, unplug connections, slide metal frame around

Transformer and insert into Control Box. Screw into place to secure. 

6. Reattach back panel onto the Control Box and screw into place.

Problem: Table Does Not Raise or Lower After Table is Turned On.
Possible Causes: Transmission Cable or Shaft May Be Defective.

To Check Transmission Cable (B) (see Diagram 2):

1. Unscrew black protective cover from wire connecting the 

Transmission Shaft to the Control Box (located along side of table base).

2. Inspect wire for any damage.
3. Leave protective cover off until you finish checking Transmission Shaft.

To Replace Transmission Shaft (C) (see Diagram 3):

1. Using a wrench, loosen and remove the small bolt holding the Transmission

Shaft to the bottom of the table base (C-1).

2. Loosen and remove the large bolt holding the Transmission Shaft to the top 

of the table base (C-2).

3. Using either an allen wrench on the bolt end or an open end wrench on the

nut, loosen and remove the bolt to disconnect end of Transmission Shaft (C-3).

4. Disconnect plug ends from Transformer socket. Remove Transmission shaft.
5. Slide new Transmission shaft into place and reattach bolts. 

Note: Fasteners on end should have longer fastener on bottom and shorter
on top to correctly position shaft (C-4).

6. Run wire along table frame and reattach protective metal cover to secure. 

Make sure all excess cable length is left on the end near the Control Box.

7. Snap plastic guard over wire and insert guard in slot in side of Control Box.

Exposed wire should be snug to box. 

8. Attach cable plugs into Transformer socket and place excess wire into

Control Box.

9. Reattach bottom panel on Control Box and screw into place to secure.

A1

A2

A3

Diagram 1

Diagram 2

C1

C2

C3

C4

Diagram 3

Page 2 of 3

TP15140 elec. grooming table inst  1/12/06  11:32 AM  Page 2

Reviews: