![Mastech CSIMS8264 Operator'S Instruction Manual Download Page 13](http://html1.mh-extra.com/html/mastech/csims8264/csims8264_operators-instruction-manual_1736104013.webp)
Range Resolution
Accuracy
200
Ω
0.1
Ω
±
(0.8% of rdg+3 digits)
2k
Ω
1
Ω
20k
Ω
10
Ω
200k
Ω
100
Ω
2M
Ω
1k
Ω
±
(0.8% of rdg+1 digit)
20M
Ω
10k
Ω
±
(1.0% of rdg+2 digits)
200M
Ω
0.1M
Ω
±[
5.0% of (rdg-10digits) +10 digits]
Overload protection: 380V dc or 380Vac rms.
Open Circuit Voltage: Less than 700mV.
4.2.5 Diode
Range
Resolution
Function
1mV
Display read approx. forward
voltage of diode
Forward DC Current: approx. 1mA
Reversed DC Voltage: approx. 2.8V
Overload protection: 380Vdc or 380Vac rms.
4.2.6 Audible continuity
Range
Continuity beeper
≤
30
Ω
Open circuit voltage: Less than 700mV.
Overload protection: 380Vdc or 380Vac rms.
4.2.7 Transistor
Range
Description Test
Condition
hFE
Display read approx.
HFE value (0-1000) of
transistor under test (all
type).
Base Current approx.
10
μ
A, Vce approx.
2.8V.
4.2.8 Temperature
Range Resolution
Accuracy
-20
℃
~0
℃
±
(5.0% of rdg+4 digits)
1
℃
~400
℃
±
(1.0% of rdg+3 digits)
401
℃
~1000
℃
1
℃
±
2.0% of rdg
20
Overload protection: Resettable fuse (F200mA/250V).
* Temperature specifications do not include thermocouple
errors.
4.2.9 Capacitance
Range Resolution
Accuracy
2nF 1pF
20nF 10pF
200nF 0.1nF
2
μ
F 1nF
20
μ
F 10nF
±
(4.0% of rdg+3 digits)
Overload protection: Resettable fuse (F200mA/250V).
4.2.10 DC Current
Range Resolution
Accuracy
20mA 10
μ
A
200mA 0.1mA
±
(1.5% of rdg+1 digit)
10A 10mA
±
(2.0% of rdg+5 digits)
Overload protection: Resettable fuse (F200mA/250V). 10A
range unfused
Max. input current: 200mA dc or 200mA ac rms for mA range,
10A dc or 10A ac rms for 10A ranges.
For measurements>5A, 4 minutes maximum ON to measure
10 minutes OFF; Above 10A unspecified.
4.2.11 AC Current
Range Resolution
Accuracy
2mA 1
μ
A
±
(1.0% of rdg+3 digits)
200mA 0.1mA
±
(1.8% of rdg+3 digits)
10A 10mA
±
(3.0% of rdg+7 digits)
Overload protection: Resettable fuse (F200mA/250V). 10A
range unfused.
Max. input current: 200mA dc or 200mA ac rms for mA range,
10A dc or 10A ac rms for 10A ranges.
For measurements>5A, 4 minutes maximum ON to measure
10 minutes OFF; Above 10A unspecified.
Frequency Range: 40Hz-400Hz
Response: Average, calibrated in rms of sine wave
21
5. MAINTENANCE