Masimo RD SET Adapter Cable Series Manual Download Page 3

3

9147A-eIFU-0116

en 

RD SET™

Adapter Cable Series

DIRECTIONS FOR USE

Reusable

LATEX

PCX-2108A

02/13

  Not made with natural rubber latex

Non-Sterile

INDICATIONS

The RD SET™ Adapter Cable Series have the same indications for use as the applicable sensors. Please see the compatible 

sensor’s directions for use for indications for use and prescribed information.

DESCRIPTION

The RD SET Adapter cables allow Masimo RD SET sensors to be used on PC or M-LNC and RC patient cables on Masimo® SET® 

or Masimo rainbow® SET devices.

RD SET Adapter Cable Series consist of the following:

•  RD to PC Adapter Cable
•  RD to M-LNC Adapter Cable

The RD SET Adapter Cable Series have been verified using Masimo SET or Masimo rainbow SET technology.

WARNING:

 Masimo sensors and cables are designed for use with devices containing Masimo rainbow SET oximetry or 

licensed to use Masimo sensors.

WARNINGS

•  Always refer to the oximeter module operator’s manual for complete instructions or additional instructions.

•  Ensure the cable is physically intact, with no broken or frayed wires or damaged parts. Visually inspect the cable and 

discard if cracks or discoloration are found.

•  All sensors and cables are designed for use with specific monitors. Verify the compatibility of the monitor, cable and 

sensor before use, otherwise degraded performance and/or patient injury  can result.

•  Carefully route patient cable to reduce the possibility of patient entanglement or strangulation.

•  Failure to properly connect the sensor or the oximeter module to the cable will result in intermittent readings, 

inaccurate results, or no reading.

•  To avoid damage to the cable, always hold it by the connector rather than the cable when connecting or disconnecting 

either end.

•  To prevent damage, do not soak or immerse the sensor or cable in any liquid solution. Do not attempt to sterilize the 

cable or sensor.

•  Do not attempt to reprocess, recondition or recycle Masimo sensors or patient cables as these processes may damage 

the electrical components, potentially leading to patient harm.

INSTRUCTIONS 

A) Attaching the RD SET Adapter Cable to a Patient Cable

 RD 

to PC Adapter Cable

1.  Refer to 

Fig. 1. 

Orient the connecting tab so that the “shiny” contacts are facing up and mate to the patient cable.

2.  Insert the connector into the patient cable connector until there is a tactile or audible click of connection.
3.  Gently tug on the connectors to ensure a positive contact. 

 RD 

to M-LNC Adapter Cable

1.  Refer to 

Fig. 2. 

Properly orient and insert the connector completely into the patient cable connector.

2.  Refer to 

Fig. 3. 

Completely close the protective latch.

B) Disconnecting the RD SET Adapter Cable from a Patient Cable

 

RD to PC Adapter Cable

1.  Refer to 

Fig. 4. 

Place thumb and index finger on the buttons on either side of the connector.

2.  Press firmly on the buttons and pull to remove.

 RD 

to a M-LNC Adapter Cable

1.  Refer to 

Fig. 5. 

Lift the protective latch to gain access to the connector.

2.  Refer to 

Fig. 6. 

Pull firmly on the connector to remove from the patient cable. To avoid damage, pull on the connector, not 

the cable.

CLEANING

1.  Remove the sensor from the patient and disconnect it from the adapter cable .

2.  Clean the surface of the cable by wiping it with a 70% isopropyl alcohol pad .

3.  Wipe all surfaces of the cable 

4.  Saturate another cloth or gauze pad with sterile or distilled water and wipe all surfaces of the cable .

5.  Dry the cable by wiping all surfaces with a clean cloth or dry gauze pad.

CAUTION:

•  Do not immerse the connector on the cable in any liquid solution.

•  Do not sterilize by irradiation, steam, autoclave, or ethylene oxide.

•  Do not clean with chemicals not approved above.