Masimo Radical 7 Operator'S Manual Download Page 3

3

  Changes or modifications to the wireless radio feature whether intentional or 

unintentional are prohibited without written approval from Masimo Corporation.

SAFETY INFORMATION, WARNINGS, CAUTIONS AND NOTES (CONTINUED) 

■ 

The Radical-7 (device with optional radio) wirelessly transmits real-time sensor connectivity 
status, indicating a connect and/or disconnect state. If the device is in a failure mode then 
the radio power is disabled and an error message is indicated on the device display.  The 
device does not have a powered state where no information is transmitted.

■ 

In accordance with FCC requirements, the Radical-7 (device with optional radio) 
must be placed greater then 20 cm from the patient’s head or torso.

■ 

In accordance with FCC requirements, radio accessories on the Radical-7 (device 
with optional radio) cannot be attached directly to the patient using any accessory 
containing metal components.

■ 

In accordance with international telecommunication requirements, the frequency 
band of 5,150 MHz to 5,250 MHz is only for indoor usage to reduce potential for 
harmful interference to co-channel mobile satellite systems.

SECTION 7:  SPECIFICATIONS 

Wireless Radio Specifications (if installed)

Radio Modes

8

802.11 a/b/g

Encryption

64/128-bit WEP, Dynamic WEP, WPA-TKIP, WPA2-AES

Authentication

Open System, Shared Key, Pre-Shared Key (PSK), 

802.1X: LEAP, PEAP, TTLS, TLS, EAP-FAST

Compliance

USA

FCC ID: VKF-RAD7CPL

FCC parts 15.207, 15.209, 15.247 and 15.407

Canada

IC: 7362A-RAD7CPL

RSS-210

Europe

EN 300 328
EN 301 893

EN 301 489-17

Bluetooth Specifications (if installed)

RF output power (conducted 
- excluding antenna gain)

Class 2 , max 4dBm (2.5mW)

Output frequency

2.402 - 2.480 GHz, ISM band

Referenced QDIDs

B016013  combined  with B015102

Modulation

FHSS (Frequency hopping)

Compliance

USA

FCC ID: PVH0925

Canada

IC:  5325A-0925

Europe

EN 300 328

EN 301 489-1

EN 301 489-17

EN 61000-6-2