![Masimo Rad-97 Operator'S Manual Download Page 85](http://html1.mh-extra.com/html/masimo/rad-97/rad-97_operators-manual_1735240085.webp)
Rad-97
Chapter 4: Operation
www.masimo.com
83
Masimo
MAP Alarms
From the
Mean Arterial Pressure
screen, touch
Alarms
, and then change any of the following
options:
Options Description
Alarm
Priority
Factory
Default
Settings
User Configurable
Settings
High
Limit
High Limit is the upper
threshold that triggers an
alarm.
Medium
120
28-219 in steps of 1,
or Off
When set to Off,
alarm is disabled
Low
Limit
Low Limit is the lower
threshold that triggers an
alarm.
Medium
50
Off, or 27-218 in
steps of 1
When set to Off,
alarm is disabled
Pulse Rate (PR)
From the
Pulse Rate Settings
screen, access the following screens:
Pulse Rate Alarms
on page 83
About Parameter Information
on page 78
Pulse Rate Alarms
From the
Pulse Rate Settings
screen, touch
Alarms
, and then change any of the following
options:
Options Description
Alarm
Priority
Factory
Default
Settings
User Configurable
Settings
High
Limit
High Limit is the upper
threshold that triggers an
alarm.
Medium
120
40-210 in steps of 5,
or Off
When set to Off,
alarm is disabled
Low
Limit
Low Limit is the lower
threshold that triggers an
alarm.
Medium
50
Off, or 35-215 in
steps of 5
When set to Off,
alarm is disabled