Chapter 4: Operation
www.masimo.com
76
Masimo
SpCO Settings
From the
SpCO Setting
s screen, access the following screens:
SpCO Alarms
on page 76
In Vivo for SpCO
on page 76
Trends
on page 97
About Parameter Information
on page 78
SpCO Alarms
From the
SpCO Settings
screen, access the following screens:
Options
Description
Alarm
Priority
Factory
Default
Settings
User Configurable
Settings
High Limit High Limit is the upper
threshold that triggers an
alarm.
High
10
2% to 98%, in steps
of 1%, or Off
When set to Off,
alarm is disabled
Low Limit Low Limit is the lower
threshold that triggers an
alarm.
Medium
Off
Off or 1% to 97%, in
steps of 1%
When set to Off,
alarm is disabled
Silence
Duration
Sets the amount of time
that the alarm is silenced.
NA
2 minutes
30 seconds, 1, 2, or 5
minutes
In Vivo for SpCO
In Vivo is currently not available in the U.S.A. and territories relying on FDA market
clearance.
From the
In Vivo
screen, access the following screens:
Options
Description
Factory Default
Settings
User Configurable
Settings
Enabled
See
In Vivo Adjustment
Overview
on page 61.
Off
On or Off