Rad-97
Chapter 4: Operation
www.masimo.com
68
Masimo
RRa Settings
RRa is active when the following conditions are all met:
•
RRa is installed on the Rad-97.
•
A dual rainbow cable is connected.
•
An acoustic sensor is connected.
Note:
See the Directions for Use provided with the acoustic sensor.
When using an acoustic sensor, Respiration Rate (RR) is determined by the acoustic (RRa)
signal. See
rainbow Acoustic Monitoring™ (RAM™)
on page 31. When the respiratory rate is
determined by the acoustic signal, the
Main Screen
labels respiratory rate as
RRa
, as shown
below.
From the
RR Settings
screen, access any of the following screens:
RRa Alarms
on page 68
Additional Settings for RRa
on page 69
About Parameter Information
on page 78
RRa Alarms
From the
Alarms
screen, change any of the following options:
Options
Description
Alarm
Priority
Factory
Default
Settings
User Configurable
Settings
High Limit
High Limit is the upper
threshold that triggers an
alarm.
High
30 breaths
per minute
6 to 69 breaths per
minute in steps of 1
breath per minute, or
Off
Low Limit
Low Limit is the lower
threshold that triggers an
alarm.
High
6 breaths
per minute
Off, or 5 to 68 breaths
per minute in steps of 1
breath per minute
Silence
Duration
Sets the amount of time
that the alarm is silenced.
NA
2 minutes
30 seconds or 1, 2 or 5
minutes