Signal Extraction
Pulse Oximeters
O P E R A T O R ’ S
M A N U A L
Page 1: ...Signal Extraction Pulse Oximeters OPERATOR S MANUAL...
Page 2: ...9 297 7001 www masimo com EU Authorized Representative for Masimo Corporation MDSS GmbH Schiffgraben 41 30175 Hannover Germany Tel 49 511 62 62 86 30 Fax 49 511 62 62 86 33 MEDICAL ELECTRICAL EQUIPMEN...
Page 3: ...If sensor is exposed to the irradiation the reading might be inaccurate or the unit might read zero for the duration of the active irradiation period Always remove the sensor from the patient and com...
Page 4: ...ory default settings 4 10 Setup Menu Level 3 Clear trend 4 10 t a b l e o f c o n t e n t s iv Rad 8 Signal Extraction Pulse Oximeter Operator s Manual Rad 8 Signal Extraction Pulse Oximeter Operator...
Page 5: ...the Rad 8 pulse oximeter appears before this introduction Other important safety information is located throughout the manual where appropriate Read the entire safety information section before you o...
Page 6: ...ort 8 hours Internal battery life with fully charged battery Serial output port Display capability on Philips Agilent monitor through Philips VueLink function RadNet TM and RadLink capability INDICATI...
Page 7: ...able to adapt to the varying physi ologic signals and or noise and separate them by looking at the whole signal and breaking it down to its fundamental components The Masimo SET signal processing alg...
Page 8: ...he bottom of the Rad 8 vertical Rad 8 family offers full Masimo SET technology in a small compact device Rad 8 family supports the full line of Masimo sensors and patient cables see Section 8 Sensors...
Page 9: ...rtical LED bar rises and falls with the pulse where the height of the bar indicates the quality of the signal 9 Saturation Display The functional arterial hemoglobin oxygen saturation is displayed in...
Page 10: ...er or packaging and are defined below SYMBOLS RS 232 Equipotential Ground Terminal Caution consult accompanying documents Nurse Call Interface WEEE compliant Defibrillation Proof see front panel Mark...
Page 11: ...er source CAUTION DO NOT CONNECT THE RAD 8 PULSE OXIMETER TO AN AC OUTLET CONTROLLED BY A SWITCH Verify the AC power voltage and frequency before use Verify that the power source can provide adequate...
Page 12: ...e front panel display is free of alarm and system failure messages see Section 5 Alarms and Messages 9 Verify the display shows the following Mode setting Standard Std or Sleep SLP or Home Hnn SpO2 Lo...
Page 13: ...DISABLED IN THIS MODE If the unit is connected to a RadNet system there will be no communication with RadNet in this mode 11 Verify that the patient alarms are functional by setting the high and low...
Page 14: ...ity of the readings over time is affected by the averaging mode being used The longer the averaging time the more stable the readings tend to become This is due to a dampened response as the signal is...
Page 15: ...rovides a visual indicator of the plethysmogram signal quality and an alert when the displayed SpO2 values are not based on adequate signal quality The signal quality indicator displayed on the Rad 8...
Page 16: ...ound and can be silenced for 120 seconds default by pressing the Alarm Silence Button Refer to Setup Menu Level 1 in this section to change setting If a low battery condition occurs while not monitori...
Page 17: ...rend memory while the device is turned on and the trend data remains in memory until the memory fills up or cleared by the user NOTE It is recommended that you clear the trend prior to performing a ne...
Page 18: ...ing choose the preferred settings by stepping through menu options Refer to Section 5 Output 2 Connect one end of the VueLink cable to the Serial Output connector on the back of the Rad 8 3 Connect t...
Page 19: ...e front panel of the oximeter The following sub sections describe each menu item in more detail The oximeter has options that allow user configuration to suit specific needs SPECIAL MENU STANDARD HOME...
Page 20: ...Sensor 100 reserved 200 reserved 400 Low Signal IQ 800 Masimo SET This flag means the algorithm is running is full SET mode It requires a SET sensor and needs to acquire some clean data for this flag...
Page 21: ...THE LIMITS EACH TIME THE PULSE OXIMETER IS USED An audible alarm and a flashing alarm status indicator will occur when an alarm limit is exceeded for greater than five seconds See section 4 Alarm Fea...
Page 22: ...taneously Return for service 5 a l a r m s a n d m e s s a g e s MESSAGES The Rad 8 will indicate other data or system errors Message conditions for the Rad 8 follow DISPLAY TYPE SOLUTION LEDS FLASH H...
Page 23: ...s Pulse Rate Accuracy Pulse rate 25 240 bpm No Motion1 Adults Pediatrics Neonate 3 digits Motion2 Adults Pediatrics Neonate 5 digits Low Perfusion3 Adults Pediatrics Neonates 3 digits Resolution Satur...
Page 24: ...lass B Equipment Classification IEC 60601 1 UL 60601 1 IEC 60601 1 1 Type of Protection Class 1 on AC power Internally powered on battery power Degree of Protection Patient Cable Type BF Applied Part...
Page 25: ...f monitoring For more information refer to the following table or contact your Sales Representative Use only Masimo SET sensors and sensor cables Select an appropriate sensor apply it as directed and...
Page 26: ...o L 3 kg 3 3 3 bpm 5 bpm 3 3 bpm 40 kg 2 3 3 bpm 5 bpm 2 3 bpm LNCS NeoPt L 1 kg 3 3 3 bpm 5 bpm 3 3 bpm SENSOR APPLICATION INSTRUCTIONS Unless indicated otherwise in the directions for use reposition...
Page 27: ...OLUTION DONOTSTER ILIZE ANY MASIMO SENSOR BY IRRADIATION STEAM OR ETHYLENE OXIDE M a s i m o S E T p a t i e n t c a b l e s Reusable patient cables of various lengths are available Only use appropria...
Page 28: ...ot allow liquids to enter the interior of the instrument Glutaraldehyde Solution Ammonium Chloride Wipe 10 Chlorine bleach in H2O 70 Isopropyl alcohol CAUTIONS DO NOT AUTOCLAVE PRESSURE STERILIZE OR G...
Page 29: ...ibed tests discontinue its use and correct the problem before returning the unit back to the user Before performing the following tests verify unit is connected to AC power Also disconnect any patient...
Page 30: ...part of the price paid for this product MASIMO grants to Purchaser a nonexclusive nontransferable license without right to sublicense to use the copy of the incor porated software firmware and documen...
Page 31: ...vative works based on the written materials without the prior written consent of MASIMO 6 Transfer Restrictions The software firmware is licensed to the Purchaser and may not be transferred to anyone...
Page 32: ...tion on both the sensors and monitors to ensure accurate pulse oximetry when needed most 2007 Masimo Corporation Masimo Signal IQ Discrete SaturationTransform DST Satshare SET FastSat Rad 8 Radlink Fa...