background image

Doc.Ref. No. m61C/QG/501         Issue No.:00 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Page 2 of 2 

 

 

Decrement 

Key 

 

 

 

Decrement the 

Value of any 
Parameter in edit 

mode. 

 

Shuffle to previous 
parameter in the 

Sub Menu. 

 

Shows ambient 

value for T/C 

Input in RUN 
mode. 

Enter Key 

(Menu Key) 

 

 

It is used to enter 
in the sub menu. 

 

It’s used to enter 
in the edit mode. 

 

It is also used to 

save the 
parameters to 

nonvolatile 
memory, when 

user setting a 

proper data by 
Increment and 

decrement key for 
parameter 

configuration. 

 

PV(Process 

Value)Display 

PV 

 

 

4 digit 0.80 inch 
RED Display 

 

Display process 
value. 

 

Display parameter 

name when user 
set parameter. 

 

Display Parameter 
Value when in Edit 

mode. 

 

Display error 
message when an 

error occurs. 
 

 

PARAMETER SETTINGS 

 

Parameter  

(PV display) 

Setting Name  
& Description 

Default  

Value 

Symbol 

Name 

 

INPT 

(

inpt

) 

 
 

INPUT 

Type 

Set PV Input 
Type 

tC j 

tC t 

tC 

tC r 

tC s 

rtd1 

rtd.1 

0-5v 

1-5v 

0-

10 

(ReferTable1.1) 

 

RTD.1 

 

ZERO 

(

zero

) 

 

Zero 

Automatically 
change to the 
Input Lower 
Range with 
changing of 
Input Type 
(ReferTable1.1)
Can be set to 
any value within 
the Input Range 
& less the SPAN 
Value. 

 

-199.9 

 

 

SPAN 

(

spaN

) 

 

Span 

Automatically 
change to the 
Input Higher 
Range with 
changing of 
Input Type 
(ReferTable1.1) 
Can be set to 
any value within 
the Input Range 
& greater the 
ZERO Value. 

 

850 

 

 
 

*DP 

(

dp

) 

 
 
 

Decimal 

Point 

Set position of 
Decimal Point 
on Display.              

0

0.0

0.00

0.000 

0:0    
1:0.0       
2:0.00        
3:0.000 

BRHT 

(

brHt

 

Brightn-

ess 

Adjust 
Brightness of 
the 7-segment 
Display.            
10 to 100 

50 

PASS 

(

pass

Password 

Set Device 
Password            
0 to 99 

VERS 

(

vers

Version 

Shows the 
Version of the 
Current 
Firmware 

*parameter shows only if input type is linear. 

Cal

 Mode

 

(PV display) 

Setting Name & 

Description

 

Symbol 

Name 

**AMB 

(

Amb

Ambient 

Ambient Adjustment 

ZERO 

(

zero

Calibration 

Zero 

Calibration  Zero  for  PV 
Input(PV/SV  Display  : 

PV) 

SPAN 

(

spaN

Calibration 

Span 

Calibration  Span  for  PV 
Input(PV/SV  Display  : 

PV) 

**parameter showsonly if input type is TC. 

MENU LAYOUT FOR 408-M 

 

 

 

Examples: 

How to change Set Point 

 

How to change Input Type 

 

 

For operation manual please visit www.masibus.com 
Specifications are subject to change without notice due to  
Continuous improvements. 
Masibus Automation And Instrumentation Pvt. Ltd. 
B-30, GIDC Electronics Estate, Sector-25, Gandhinagar- 
382044, Gujarat, India. 
Tel:+91 79 23287275-79     Fax: +91 79 23287281 
Web:www.masibus.comEmail:[email protected] 

 

 