background image

4

•  Products with a welded plastic housing are 

not repairable. Please contact your supplier for 
replacement part.

•  This product contains hazardous voltages and 

there are no user replaceable parts inside the 
product. Never attempt to remove the cover.

•  WARNING: No modifi cation of this equipment is 

allowed. Any repair/service should be carried out 
by qualifi ed personnel who may get assistance by 
contacting the manufacturer or the manufacturer’s 
agent.

•  Products specifi ed to have automatic polarity 

protection must be switched off if a battery is 
connected with reverse polarity. The protection will 
be automatically reset when the polarity has been 
corrected.

•  In chargers specifi ed to have a replaceable fuse as 

polarity protection the fuse must be replaced if the 
battery has been connected with reverse polarity. 
When replacing the fuse; a fuse of the same type 
and rating must be used.

•  If the product is specifi ed to comply with the 

standard for Medical Electrical Equipment 
(standards based on IEC60601-1) it complies 
with some of the requirements for medical 
electrical equipment and may be used in medical 
applications and hospital environments.

•  The product must not be used in the vicinity 

of fl ammable anesthetic gases or in other 
environments with fl ammable or explosive 
atmosphere. 

•  If the product is specifi ed to comply with the 

standard for Medical Electrical Equipment 
for Home Healthcare Environment (standard 
IEC60601-1-11) it may be used in medical 
applications used in a home healthcare 
environment.

  NOTE: Products relying on safety earth for 

protection (Class I) may not be used in home 
healthcare environment unless they are 
permanently wired to the building installation:               

  Installation must only be carried out by qualifi ed 

service personnel, following the below instructions:
               

-  The protective earth conductor must be

min. 0.75 mm

2

.           

-  Connect the protective earth conductor to the 

external protective earthing system.   

-  Verify that the protective earth terminal used is 

connected to the external protective earthing 
system.             

-  Verify the integrity of the external protective 

earthing system.

•  This product converts the mains voltage to 

a safety extra low voltage. 

  The output from products applying with 2MOPP 

insulation ( model names followed by ”P”) may 
be treated as Applied Part Type B or Type BF 
according to standard EN/IEC 60601-1 and 
may come in physical contact with a patient.The 
housing of the product shall not be allowed to 
contact the patient .

•  This product must be operated in an environment 

within temperature range +5 to +40°C, humidity 
15 - 93 % RH and atmospheric pressure 70 - 
106 kPa (700 - 1060 hPa).

•  Expected service life of this product and 

accessories delivered with this product is three (3) 
years, if operated as indicated above. However; the 
guarantee times indicated in document ”TERMS 
OF SALES AND DELIVERY FOR MASCOT AS” 
apply (available at www.mascot.com).

•  Environmental parameters during transport and 

storage between uses: temperature range -25 
to +85 °C, humidity 15 - 93% RH NC and 
atmospheric pressure 70 - 106 kPa (700 - 1060 
hPa). 

•  If stored for longer periods of time the 

environmental parameters should be within 
the temperature range +5 to +35°C, humidity 
range 10 - 75% RH NC and atmospheric pressure 
70 - 106 kPa (700 - 1060 hPa) to maintain the 
products expected service life.

•  Expected shelf life of this product is one (1) year, if 

stored as indicated above.

Summary of Contents for 205140

Page 1: ...w mascot no downloads usermanuals Bruksanvisning Käyttöohjeet Bedienungsanleitung Mode d emploi Manual de instrucciones Istruzioni per l uso Doc no 5140C Part No 205140 04 12 2017 MASCOT ELECTRONICS AS P O Box 177 N 1601 Fredrikstad NORWAY Phone 47 69 36 43 00 Telefax 47 69 36 43 01 E mail sales mascot no Web www mascot no ...

Page 2: ...RIC SHOCK READ THROUGH THESE INSTRUCTIONS PRIOR TO USING THE PRODUCT CAREFULLY FOLLOW THESE INSTRUCTIONS WHEN USING THE PRODUCT RETAIN THESE INSTRUCTIONS FOR FUTURE REFERENCE compound and is dust tight and protected against the effects of temporary immersion in water according to standard EN IEC 60529 but must not be immersed in water for longer periods of time Products marked with the double squa...

Page 3: ...emoval of the products mains supply should an operational error occur during use If the product has a detachable mains cord the appliance coupler may be used as a disconnect device The product is switched on by inserting the mains plug into the mains socket and switched off by disconnecting the mains plug from the mains socket The product may be connected to an IT type mains supply For use in U S ...

Page 4: ...may not be used in home healthcare environment unless they are permanently wired to the building installation Installation must only be carried out by qualified service personnel following the below instructions The protective earth conductor must be min 0 75 mm2 Connect the protective earth conductor to the external protective earthing system Verify that the protective earth terminal used is conn...

Page 5: ...rs result in a fuse blowing leaving the charger useless When charging batteries other than vehicle batteries it is recommended that you connect the charger to the mains before connection is made to the battery This will reduce the sparks Cautions to observe prior to charging Lead Acid batteries This product complies with the requirements to electromagnetic compatibility for medical electric equipm...

Page 6: ...ore removing battery connections If the charger is equipped with battery clips first disconnect the clip from the battery s positive pole then the other from the negative battery pole vehicle chassis in this order The recommended minimum battery capacity for which the specific charger can be used vary from battery to battery Some do not have a capacity restriction at the specified boost voltage le...

Page 7: ...50V 13A BS 1363 AUS 250V 10A AS NZS 3112 Mains Cord Set is available on request if you wish your product to be DeskTop How to connect exchangeable DC output plugs 1 To connect for desired polarity both plug ends are clearly marked 2 When connected the female plug is also marked on each side to identify plug polarity 3 Shows the center polarity of the plug 3 12 3 1 2 ...

Page 8: ... The LED indication on the charger is ORANGE The battery is charged to its full capacity at the end of this step STEP 3 FLOAT CHARGE The LED indication on the charger is GREEN and the battery is fully charged The charger is in standby mode The charge voltage is at standby level and the charger may remain connected to the battery The charger will return to boost charging if the battery is used A lo...

Page 9: ...dicator charges to yellow The charger stays in this mode until the charge current decreases to charge termination level The battery is charged to its full capacity at the end of this step STEP 3 FLOAT CHARGE The LED indication on the charger is GREEN and the battery is fully charged The charger is in standby mode The charge voltage is at standby level and the charger may remain connected to the ba...

Page 10: ...r mode indicated by the YELLOW LED and will remain in this mode until time interval is completed The battery is charged to its full capacity at the end of this step STEP 3 FLOAT CHARGE The LED indication on the charger is GREEN and the battery is fully charged The charger is in standby mode The charge voltage is at standby level which means that the charger can continue to be connected to the batt...

Page 11: ... charger is in standby mode The charge voltage is at standby level and the charger may remain connected to the battery The charger will return to boost charge if the battery is used BATTERY NOT CONNECTED INDICATIONS Battery not connected is indicated by FLASHING GREEN Yellow Yellow Green INDICATOR FLASHING YELLOW INDICATOR GREEN Charge Voltage Charge Current INDICATOR YELLOW Charge Voltage Battery...

Page 12: ...y charged The charger is in standby mode The charge voltage is at standby level and the charger may remain connected to the battery The charger will return to boost charge if the battery is used BATTERY NOT CONNECTED INDICATIONS Battery not connected is indicated by FLASHING GREEN In this mode charger will apply short pulses attempting to wake up deeply discharged batteries ERROR INDICATIONS 2 red...

Page 13: ...A 44 1V 44 1V 0 4A 44 1V 0 4A 41 1V 250mA 4Ah 12 5Ah 0 6A 58 8V 58 8V 0 25A 58 8V 0 25A 54 8V 100mA 3Ah 5Ah Input voltage Charge LED indicator 6V 12V 24V 36V 48V 2240 P 2241 P 100 240Vac 50 60Hz max 0 35A Orange CC ch Orange CV ch Green Float ch Rec batt capacity 1 3A 7 35V 7 35V 250mA 6 85V 250mA 6 5Ah 12 5Ah 1A 14 7V 14 7V 250mA 13 7V 250mA 5Ah 12 5Ah 0 56A 29 4V 29 4V 250mA 27 4V 250mA 2 8Ah 12...

Page 14: ... 284X P 314X P 220 240Vac 50 60Hz Orange CC ch Orange CV ch Yellow CV ch Green Float ch Rec batt capacity 8 5A 7 35V 7 35V 4 25A 7 35V 4 25A 6 85V 1 6A 42 5Ah 80Ah 7A 14 7V 14 7V 3 5A 14 7V 3 5A 13 7V 1 6A 35Ah 80Ah 3 5A 29 4V 29 4V 1 7A 29 4V 1 7A 27 4V 0 8A 17 5Ah 40Ah 2 3A 44 1V 44 1V 1 1A 44 1V 1 1A 41 1V 0 5A 11 5 25Ah 1 7A 58 8V 58 8V 0 9A 58 8V 0 9A 54 8V 0 4A 8 5Ah 20Ah 324X P 110 120Vac 2...

Page 15: ...ch Yellow Timer CV ch Green Float ch Rec batt capacity N A 4A 14 7V 14 7V 4A 2h 13 8V 20Ah 200Ah 2A 29 5V 29 5V 2A 2h 27 6V 10Ah 100Ah 1 4A 44 1V 44 1V 1 4A 2h 41 4V 7Ah 70Ah 1A 58 8V 58 8V 1A 4h 54 8V 5Ah 50Ah 9840 220 240Vac 50 60Hz Orange CC ch Yellow Timer CV ch Green Float ch Rec batt capacity 5A 7 35V 7 35V 5A 2h 6 85V 25Ah 250Ah 5A 14 7V 14 7V 5A 2h 13 7V 25Ah 250Ah 2 5A 29 4V 29 4V 2A 2h 2...

Page 16: ...A 29 4V 29 4V 10A 4h 27 4V 50Ah 500Ah 6 7A 44 1V 44 1V 6 7A 4h 41 1V 33 5Ah 335Ah 5A 58 8V 58 8V 5A 4h 54 8V 25Ah 250Ah 2640 per channel 220 240Vac 50 60Hz Orange CC ch Yellow Timer CV ch Green Float ch Rec batt capacity 10A 7 35V 7 35V 10A 4h 6 85V 50Ah 500Ah 10A 14 7V 14 7V 10A 4h 13 7V 50Ah 500Ah 5A 29 4V 29 4V 5A 4h 27 4V 25Ah 250Ah 3 3A 44 1V 44 1V 5A 4h 41 1V 16 5Ah 165Ah 2 5A 58 8V 58 8V 2 ...

Page 17: ...0Vac 50 60Hz Yellow CC ch Flash Yellow CV ch Green Float ch Rec batt capacity 5A 7 35V 7 35V 1 5 5A 4h 6 85V 0 5A 25Ah 250Ah 5A 14 7V 14 7V 1 5 5A 4h 13 7V 0 5A 25Ah 250Ah 2 5A 29 4V 29 4V 0 6 2 5A 4h 27 6V 0 2 5A 12 5Ah 125Ah 1 7A 44 1V 44 1V 0 4 1 7A 4h 41 1V 0 1 6A 8 5Ah 85Ah 1 2A 58 8V 58 8V 0 3 1 2A 4h 54 8V 0 1 2A 6Ah 60Ah 2245 220 240Vac 115Vac 50 60Hz Yellow CC ch Flash Yellow CV ch Green ...

Page 18: ...loat ch Rec batt capacity 2 7A 7 35V 7 35V 0 5 2 7A 4h 6 85V 0 2 7A 13 5Ah 135Ah 2A 14 7V 14 7V 0 5 2A 4h 13 7V 0 2A 10Ah 100Ah 1A 29 4V 29 4V 0 25 1A 4h 27 4V 0 1A 5Ah 50Ah 0 65A 44 1V 44 1V 0 15 0 65A 4h 41 1V 0 0 65A 3 25Ah 32 5Ah N A 3540 230Vac 50 60Hz Yellow CC ch Flash Yellow CV ch Green Float ch Rec batt capacity N A 20A 14 7V 14 7V 2 5 20A 4h 13 7V 0 20A 100Ah 1000Ah 10A 29 4V 29 4V 1 4 1...

Reviews: