Martini Instruments MI 415 User Manual Download Page 2

GUIDE TO DISPLAY CODES:

This prompt appears for 1
second each time the instru-
ment is turned on.

The dashes “- - -” indicates
that the meter is in a ready
state and measurement can
be  performed.

Sampling In Progress. Flash-
ing “SIP” prompt appears each
time the meter is performing a
measurement.

The presence of battery icon
on the display indicates that
the battery voltage is  getting
low and the battery needs to
be replaced.

“-BA-”, the battery is dead and
must be replaced. Once this
indication is displayed, the
meter will lock up. Change the
battery and restart the meter.

“CAL”, the meter is in calibra-
tion mode. If the CAL key is
not pressed within 6 seconds,
the meter will automatically
switch to idle mode.

“CL” appears on the lower part
of the display when the meter
is in calibration mode.

BATTERY REPLACEMENT

Battery replacement must only take place in a
non-hazardous environment.

Simply rotate the battery cover on the back of
the meter.

Detach the battery from the terminals and
attach a fresh 9V battery while paying  atten-
tion to the    correct polarity. Insert the battery
and replace the cover.

ACCESSORIES:

MI0004

Tissue for wiping cuvets
(4 pcs.)

MI0005

9V battery (1 pc.)

MI0011

10 mL glass cuvete (2pcs.)

MI0012

cap for 10 mL glass cuvet
(2 pcs.)

MI0013

Stopper for 10 mL glass
cuvet (2 pcs.)

MI515-000

Calibration standards: 0,
10, 500FNU (3 x 30 mL
dropper bottles)

WARRANTY

This instrument is warranted against defects in mate-
rials and manufacturing for a period of two years
from the date of purchase.
If during this period the repair or replacement of parts
is required, where the damage is not due to negli-
gence or erroneous operation by the user, please
return the meter to either distributor or our office in
the original packing and the repair will be free of
charge.
Damages due to accidents, misuse, tampering or
lack of prescribed maintenance are not covered.

ISTMI415  07/05

7• At the end the blinking “10.00” message

will appear on the LCD.

Insert the 10 FNU standard and press CAL
key. “SIP” will blink during measurement.

8• The blinking “500” message will appear on

the LCD.

Insert the 500 FNU standard and press CAL
key. “SIP” will blink during measurement.

9• At the end of the reading the instrument

goes directly to measurement mode. Now
the instrument is calibrated and ready to
use.

Note: 

A monthly calibration is recommended.

 The instrument can be checked with the
supplied standard solutions each day.

To check the date of the last calibration
simply press DATE/   key.

Reviews: