DC CURRENT
Ranges: 200µA, 20mA, 200mA, 10A
Resolution: 0.1µV
Accuracy: ± (1.5% rdg + 1 digit) except 10A range
± (3% rdg + 3 digits) on 10A range
Input protection: 0.5A, 250V fast blow & 10A, 600V fast blow ceramic,
fuse.
AC CURRENT (50Hz to 500Hz)
Ranges: 200µA, 20mA, 200mA, 10A
Resolution: 0.1µA
Accuracy: ± (2% rdg + 4 digits) except 10A range
± (3.5% rdg + 4 digits) on 10A range
Input protection: 0.5A, 250V fast blow & 10A, 600V fast blow ceramic,
fuse.
RESISTANCE
Ranges: 200
Ω
, 20k
Ω
, 200k
Ω
, 20M
Ω
Resolution: 100m
Ω
Accuracy: ± (1.5% rdg + 4 digits) except on 20M
Ω
range
± (3% rdg + 4 digits) for 20M
Ω
range
Open circuit voltage: 0.3V DC typical (3V DC for 200
Ω
range)
CONTINUITY
Audible tone: <100
Ω
Response time: 500mS
Overload protection: 500V DC or AC rms.
DIODE TEST
Test current: Approximately 1mA
Accuracy: ± (2% rdg + 1 digit)
Open circuit voltage: 3V DC typical
Non-contact Voltage Indicator: Operates for voltages from 70V to 480V
AC, 50-60Hz.
NOTE:
This instrument complies with class ll, overvoltage CAT lll, 600V of the
IEC1010-1 (EN61010-1) standard and pollution degree 2 in accordance
with IEC-664 indoor use.
This instrument also complies with the requirements of the following
European Community Directives: 89/336/EEC (Electromagnetic
Compatibility) and 73/23/EEC (Low Voltage) as amended by 93/68/EEC
(CE Marking).
However, electrical noise or intense electromagnetic fields in the vicinity of
the equipment may disturb the measurement circuit. Measurement
instruments will also respond to unwanted signals that may be present.
Users should exercise care and take appropriate precautions to avoid
misleading results in the presence of electromagnetic interference.