When the DTL84 is connected to an external device (PC, laptop, tablet etc), the recall 
function is disabled.

3.10 Max/Min/Avg

When this function is selected, the DTL84 simultaneously monitors and stores the maximum, 
minimum and average results. The DTL84 will continue to update and refresh the data until the 
user exits this function

To enter the function and display the maximum recorded value, press 

W

MIN

MAX
AVG

The 

MAX 

symbol will 

be displayed. 

Successive presses of 

W

MIN

MAX
AVG

will display the following in sequence:

The minimum recorded value.

 

The 

MIN

 symbol will be displayed.

The average recorded value

The 

AVG

 symbol will be displayed

Real time readings 

W

MIN

MAX
AVG

The 

MAX

MIN

 and 

AVG

 symbols flash.

To exit MAX/MIN/AVG, press 

W

MIN

MAX
AVG

 

and hold for 2 seconds.

3.11 Temperature Scale Selection

Press 

°

°

F

W

 to switch between Celsius (°C) and Fahrenheit (°F)

3.12 Temperature Measurement

The maximum temperature measurement of the DTL84 for a K-type thermocouple is 1372°C. 
The K-type thermocouple probe supplied is specified to 200 °C. Be sure that the thermocouple 
probe is adequate for the measurement being made.

Connect a thermocouple probe/s, suitable for the type of temperature measurement and 
temperature range being made, to the thermocouple input socket/s.

Set the DTL84 for the selected thermocouple type (see 4.3).
Taking all necessary safety precautions position the thermocouple/s at the surface or in the 
medium to be measured and read the measured temperature from the display.

Note:  

Repeated flexing can break the thermocouple leads. To prolong lead life, avoid sharp 
bends in the leads, especially near the connector.

3.13 Differential Temperature Measurement

Connect thermocouple probes, suitable for the type of temperature measurement and 
temperature range being made, to the thermocouple input sockets T1 and T2. 

Set the DTL84 to the T1-2 differential temperature mode (see 4.7) and the selected 
thermocouple type (see 4.3).

Taking all necessary safety precautions position the thermocouples at the surfaces or in the 
mediums to be measured.

The difference between the measured temperatures of thermocouples T1 and T2 will be 
displayed at the bottom of the LCD.

6

4. Set-up

Press 

SET

 to enter set-up. Press 

SET

 

again to exit set-up. 

Note:

  When the DTL84 is connected to the user’s chosen device (PC, laptop, tablet etc), set-up 

cannot be entered.

Press the arrow keys 

W

HOLD

REC

W

W

MIN

MAX
AVG

°

°

F

W

to select settings and to adjust parameters. 

Press 

MEM

W

 to save changes and move to the next set-up option.

4.1 Set Bluetooth ON/OFF 

Press 

W

 

HOLD

 or 

REC

W

 to turn the Bluetooth ON or OFF.

Note:

  When switching the Bluetooth function ON or when entering the standby/pairing mode, the 

blue LED in the 

SET

 

button flashes every 3 seconds. When successfully connected, the 

blue LED will flash every second. 

4.2 Select Thermocouple Type 

The DT84 can be set for the following thermocouple types: K, J, E, T, N, R and S.

Press 

W

MIN

MAX
AVG

or 

°

°

F

W

 to select the thermocouple type.

4.3 Setting the Data Recording Interval Time

Press 

W

MIN

MAX
AVG

or 

°

°

F

W

 to set the minutes and/or seconds.

Press 

W

HOLD

 or 

REC

W

 to increase/decrease the values. 

Note: 

The set interval time range is 1 second (00:01) to 60 minutes 59 seconds (60:59). 

7

8

4.4 Thermocouple Offset Adjustment

The offset adjustment allows an individual thermocouple to be optimised for the best 
measurement accuracy at a chosen reference temperature.

Press 

W

MIN

MAX
AVG

or 

°

°

F

W

 

to select the required thermocouple channel and it will flash on the LCD 

when selected.

Press 

W

HOLD

 or 

REC

W

 to increase/decrease the values.

The offset adjustment span is ± 5°C or ± 9°F

4.5 Setting the Alarm Limits (T1 only)

Press 

W

 

HOLD

 or 

REC

W

 to switch ON or OFF. 

When setting the alarm point, select ON and then press 

MEM

W

 to set the 

Hi

 and 

Lo 

limit. 

Note:

  If the alarm limits are exceeded, the LCD will flash 

ALARM

 

Hi

 or 

ALARM

 

Lo

 and the audible 

sounder will beep continuously.  
The 

Lo

 alarm setting cannot be greater than the 

Hi

 alarm setting.

9

4.6 Set T1-T2 Differential mode

Press the 

W

 

HOLD

 or 

REC

W

 buttons to switch the differential mode ON or OFF. 

Note: 

When the T1-T2 differential mode is ON, the T1-T2 measurement will be displayed at the 

bottom of the LCD. 

4.7 Set Auto Power Off Time

Press 

W

HOLD

 or 

REC

W

 to switch the auto power off function ON or OFF. The function is OFF when 

no time is selected. 

The following auto power off times can be selected - 10 or 30 minutes, 1, 2, 4, or 8 hours. The 
ON symbol will appear above the selected time.

4.8 Set System Clock

The DTL84 has a built in clock so that the data logger function can also record the date and time 
as well as the measured value. 

Press 

W

MIN

MAX
AVG

or 

°

°

F

W

 

to select the year, date and time.

Press 

W

HOLD

 or 

REC

W

 to increase/decrease the values.

Note:

  It is advisable to check the system clock settings after replacing the batteries. If 

necessary, reset the clock settings.

4

.9 Clearing Data Logger Records

Switch off the DTL84. Press and hold 

REC

W

 and then press 

 to power up the meter. Keep 

holding both buttons and 

REC, CLr

SUrE

, and 5 will appear on the LCD. Continue to hold both 

buttons until 5 counts down to 0. To abort clearing the memory release both buttons before 
the DTL84 counts down to 0. After the memory has been cleared the DTL84 will return to 
measurement mode.