Make sure the clamp meter and test leads are dry, clean and free 

from dust, grease and moisture while in use to avoid the danger 

from electric shock due to surface leakage.

Always test this unit on an appropriate proving device or a known 

good voltage source before and after using it to determine if a 

hazardous voltage exists in a circuit to be tested. 

Do not use

 the 

unit if it does not function correctly during proving.

Measuring/testing for a voltage/current that exceeds the specified 

limits of the unit may damage the unit and may expose the operator 

to a shock hazard. Always check the unit’s specified limits before 

use.

As a clamp meter or multimeter the unit must only be used on 

CAT III and CAT II installations up to 600V to earth, and within the 

operating temperature and humidity range specified.

If the removable probe tip caps are not fitted to the probes of the 

test leads, their measurement category becomes CAT II 1000V, 

and they 

must not be used

 on CAT III or CAT IV installations to 

avoid the risk of shorting high energy circuits and arc flash.

When this unit is used in combination with test leads, the 

measurement category of the combination is the lower 

measurement category of either this unit or the test leads used. 

Likewise if test lead accessories such as crocodile clips are also 

used, the measurement category will be the lowest measurement 

category in that combination.

Do not use

 if the battery compartment cover is not fitted.

CM57 instruct rev1.indd   10

11/11/2015   13:59:37

4

When using test leads, 

always

 keep your fingers behind the finger 

guard on the test lead probe.

When positioning the clamp jaws around a hazardous live conductor 

always

 keep your fingers behind the clamp safety protection barrier.

When making current measurements using the clamps, disconnect 

the test leads from the clamp meter terminals.

Cautions

Avoid severe mechanical shock or vibration and extreme temperature.

When using test leads avoid excessive stresses to the cable entry 

points at the probe and 4mm plug connector.

To avoid possible corrosion from leaking batteries, remove the 

batteries when the unit is not in use for an extended period.

2. INTRODUCTION

2.1 Inspection

Examine the shipping carton for any sign of damage. Inspect the 

unit and any accessories for damage. If there is any damage then 

consult your distributor immediately.

2.2 Description

The Martindale CM57 has a SMART function that automatically 

detects if AC or DC voltage is being measured when set to the 

voltage function. The SMART function also automatically determines 

if resistance or capacitance is being measured or if a diode is being 

tested.

CM57 instruct rev1.indd   11

11/11/2015   13:59:37

5

The Martindale CM55 has the following measurement functions:

‹

True rms AC current to 600A

‹

True rms AC voltage to 600V

‹

DC voltage to 600V

‹

Resistance to 6M

:

‹

Capacitance to 600μF

‹

Frequency to 100kHz

‹

Continuity with audible indication

‹

Diode testing

Further functions are:

‹

Non-contact voltage detector

‹

Display hold

‹

Auto power off

‹

Auto ranging

‹

Display backlight

2.3 Accessories

The CM57 comes with the following accessories:

‹

Carrying case

‹

Set of TL16 test leads

‹

2 x 1.5V AAA batteries

‹

Instructions

2.4 Battery Installation

Refer to section 4.1 (Battery Replacement) for the battery 

installation instructions. 

CM57 instruct rev1.indd   14

11/11/2015   13:59:38

6

3. OPERATION

3.1 General

If the clamp meter displays 

OL

 then the measurement limits of the 

range have been exceeded.

3.2 Low Battery Indication

If the 

 symbol is displayed, the batteries need replacing as 

measurement accuracy can no longer be guaranteed (see section 
4.1 Battery Replacement).

3.3 Description of Clamp Meter Elements

1

Clamp current sensing jaws

2

NCV sensing area

3

Clamp safety protection barrier

4

Clamp trigger

5

Rotary function selector switch

6

Backlight button

7

Frequency selector button

8

Hold button

9

Liquid crystal display

10

COM input terminal

11

Positive input terminal

Figure 1

CM57 instruct rev1.indd   15

11/11/2015   13:59:38

Summary of Contents for CM57

Page 1: ...e Updated instructions and product information are available at www martindale electric co uk 1 1 Meaning of Symbols and Markings Caution risk of danger refer to instructions Caution risk of electric shock Equipment protected by double or reinforced insulation Class II Application around and removal from hazardous live conductors is permitted CAT II Measurement Category II is applicable to test an...

Page 2: ...s Cautions Avoid severe mechanical shock or vibration and extreme temperature When using test leads avoid excessive stresses to the cable entry points at the probe and 4mm plug connector To avoid possible corrosion from leaking batteries remove the batteries when the unit is not in use for an extended period 2 INTRODUCTION 2 1 Inspection Examine the shipping carton for any sign of damage Inspect t...

Page 3: ...nd release the trigger to close the jaws Position the clamp meter so the conductor is central within the clamp jaws Read the measured ac current from the display Note Clamping around more than one conductor will result in an incorrect measurement 3 11 Voltage Measurements Connect the black test lead to the COM terminal and the red test lead to the terminal Set the rotary switch to the V position T...

Page 4: ...can be accessed by removing the screw and lifting off the cover 12 Figure 2 13 Fit 2 new 1 5V AAA alkaline batteries IEC LR03 NEDA 24A observing correct polarity Replace the battery compartment cover and screw Note Do not mix old and new batteries 4 2 Test Lead Replacement If the test leads become damaged they should be replaced The replacement test leads must have the same or better overvoltage c...

Page 5: ... an implied warranty or exclusion or limitation of incidental or consequential damages the limitations and exclusions of this warranty may not apply to every buyer If any part of any provision of this warranty is held invalid or unenforceable by a court or other decision maker of competent jurisdiction such holding will not affect the validity or enforceability of any other provision or other part...

Page 6: ... up to 2000m Pollution degree 2 SAFETY Conforms to BS EN 61010 1 BS EN 61010 2 032 CAT III 600V Class II double insulation Specification CM57 AC Smart Clamp Multimeter EMC Conforms to BS EN 61326 1 SPECIFICATION FOR TL16 TEST LEADS Maximum voltage 1000V AC DC Maximum current 10A continuous Connector 4mm banana plug with fixed shroud Environmental Temperature Operating Storage 0 C to 40 C Altitude ...

Reviews: