5. Specifications
T/C Ranges:
J, K, T, E, R, S, N, B, L, U including – 10 to 70 mV
range
RTD Ranges:
Pt 385 (100, 200, 500, 1000 ohms)
Pt 392, JIS, Ni 120, CU10, YSI 400
Ohms Ranges:
0 to 400.00 and 400.0 to 3200.0
Accuracy:
T/C Type J: ±.3°C ±10
µ
V (±0.4°C total error)
RTD PT100: ±.1°C ±.075
Ω
(±0.3°C total error)
Ohms (400): ±0.1
Ω
Ohms (3200): ±1.0
Ω
RTD IEX-range:
0.01 mA to 3 mA
RTD Frequency
Response:
10 mS – works with all pulsed XMTR’s
Oper. Temp:
-10°C to 50°C
Storage Temp:
-40°C to 60°C
Power Supply:
4 AA Alkaline Cells, 30 hours typical operating time
Size:
7.5" x 4" x 1.5"
Range & Accuracy
Range
Min
Max
Accuracy
ohms Read (low)
0.00
400.00
0.1 ohm
ohms Read (high)
400.0
1500.0
0.5 ohm
1500.0
3200.0
1.0 ohm
Range
Min
Max
Excitation Accuracy
Current
Ohms Source (low)
5.00
400.00
0.1 to 0.5 mA
0.15 ohm
5.00
400.00
0.5 to 3 mA
0.1 ohm
Ohms Source (high)
400.0
1500.0
0.05 to 0.8 mA 0.5 ohm
1500.0
3200.0
0.05 to 0.4 mA 1.0 ohm
Thermocouple Read and Source (All errors included)
Range
Min
Max
Accuracy
J Thermocouple
-200.0
0.0
0.6°C
0.0
1200.0
0.4°C
K Thermocouple
-200.0
0.0
0.8°C
0.0
1370.0
0.5°C
T Thermocouple
-200.0
0.0
0.8°C
0.0
400.0
0.5°C
E Thermocouple
-200.0
0.0
0.5°C
0.0
950.0
0.4°C
R Thermocouple
-20
0.0
2.4°C
0
500
1.7°C
500
1750
1.3°C
S Thermocouple
-20
0
2.4°C
0
500
1.7°C
500
1750
1.4°C