The battery indicator is
displayed
The
batteries
are
discharged.
Insert new batteries
Unreasonably high value is
shown for systolic or diastolic
pressure
The sleeve was held
lower than the heart
The sleeve was not fitted
properly
You moved or spoke
during the measurement
Stay in the correct posture and
measure again.
Unreasonably low value is
shown for systolic or diastolic
pressure
The sleeve was held
higher than the heart
You moved or spoke
during the measurement
Guidelines and manufacturer’s certificate concerning electromagnetic radiation
The blood pressure monitor is intended to be used in electromagnetic environments in accordance with
the specifications below. It is incumbent on the customer or the person using the blood pressure monitor
to ensure that the electromagnetic environment in which it is used complies with these specifications.
Emission test
Compliance
Guidelines for electromagnetic environment
Radio frequency
radiation CISPR 11
Group 1
Radio frequency energy is only used in the blood pressure
monitor for the internal functions in the product. The emitted
radio frequency radiation is therefore very low and the
probability of it causing interference in nearby electronic
equipment is low.
Radio frequency
radiation CISPR 11
Class B
The blood pressure monitor is suitable for use in apartments
and similar environments, directly connected to low voltage
networks for the distribution of electricity to apartment
buildings.
Overtones IEC 61000-
3-2
Not applicable
Voltage fluctuation/
flicker IEC 61000-3-3
Not applicable
Guidelines and manufacturer’s certificate concerning electromagnetic immunity
The blood pressure monitor is intended to be used in electromagnetic environments in accordance with
the specifications below. It is uncumbent on the customer or the person using the blood pressure monitor
to ensure that the electromagnetic environment in which it is used complies with these specifications.
Immunity test
Test level as per
IEC 60601
Compliance level
Guidelines for electromagnetic
environment
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV on contact
±8 kV over air gap
±6 kV on contact
±8 kV over air gap
The floor surface should consist of
wood, concrete or clinker. In rooms
with floors covered with synthetic
material the relative air humidity
should be at least 30%.