Page 11
QUI-1124
www.markes.com
6.2 Setting the agitator
[1]
Power-on the agitator.
This displays the home
screen, which shows
the current status of the
agitator. From this
screen the method
temperature, agitation
parameters and time
can be set, as
described below.
Step [1]: HiSorb Agitator home screen.
Set
temperature
Set
timer
Temperature
set point
Elapsed time
Method time
Current
temperature
Agitation speed
Start
Set
agitation
[2]
Navigate to the ‘Set temperature’ screen by
pressing the
icon.
[3]
Set the desired temperature (between 40°C
and 90°C) using the arrow buttons.
[4]
Press <ON> to activate the heaters.
NOTES
Before running the method, ensure that
sufficient time has elapsed for the vial
block to heat.
NOTES
If the HiSorb Agitator needs to be run
at room temperature, then ensure the
temperature is set to <OFF>.
[5]
Press <OK> to return to the home screen.
[6]
Navigate to the ‘Set agitation’ screen by
pressing the
icon.
Step [6]: ‘ Set agitation’
screen.
Step [2]: ‘Set temperature’
screen.