6.4 Display options
The ACTI-VOC PLUS has multi-language capability. To select the display language,
use the navigation keys to select 'Setup' from the main menu, select 'Display
Options' from the submenu screen then select Language. English is the
default language.
Available languages include: English, Spanish, French, Dutch, German, Italian
and Portuguese.
The time, date and display units can also be set from the 'Setup' menu by selecting
the appropriate function from the submenu.
7. Maintenance
7.1 Battery maintenance
ACTI-VOC PLUS uses rechargeable nickel metal hydride (NiMH) batteries that must
be fully charged and properly maintained to obtain the maximum run time. The
battery is rated at 7.2 V and has a charge time of less than 4 hours.
WARNING
Do not open the pump case, charge or replace the battery in an
explosive atmosphere.
CAUTION
All NiMH batteries lose charge when not in use. If the battery pack has
not been charged for 3–4 days, recharge it before use. NiMH batteries
stored for extended periods should be recharged every 1–2 months to
avoid complete discharge.
CAUTION
If the battery has not been charged for more than 3 weeks, it should be
reconditioned following the instructions below.
CAUTION
Ensure the ACTI-VOC PLUS is fully charged each time as partial charging
can reduce the battery capacity over time.
CAUTION
Do not short battery terminals, as this will blow the internal fuse.
NOTES
Both charger and battery pack become warm during charging.