Markes International ACTI-VOC C-LFP-01 Instructions For Use Manual Download Page 1

ACTI-VOC

QUI-1086

Version 6 

May 2021

www.markes.com

Instruction for use

P

roduct

 r

eference

:

ACTI-VOC low-flow pump kit 

C-LFP-01 

Summary of Contents for ACTI-VOC C-LFP-01

Page 1: ...ACTI VOC QUI 1086 Version 6 May 2021 www markes com Instruction for use Product Reference ACTI VOC low flow pump kit C LFP 01 ...

Page 2: ...ontact details 6 1 7 Certifications 6 1 8 EU declaration of conformity 7 2 Components 8 3 Calibration 10 4 Sampling 12 4 1 Personal workplace monitoring 12 4 2 Indoor air monitoring 13 5 Maintenance 14 5 1 Battery 14 5 2 Pump filter 15 5 3 Back pressure adapter O ring 16 5 4 6 mm o d tubes 17 6 Parts list 17 6 1 Accessories 17 6 2 Spare parts 17 7 Specifications 18 8 Contact details 19 ...

Page 3: ...tomatic compensation prevents the need for re calibration when switching between tubes during sampling minimising down time this makes it ideal for use in Markes MTS 32 multiple tube sampler part no MTS 32 ACTI VOC is compatible with industry standard thermal desorption tubes and 6 mm o d tubes and samples in the region of 20 200 mL min ACTI VOC can also be used to sample air or gas into Tedlar ba...

Page 4: ... otherwise provided by the seller or from not adhering to all federal state and local environmental and occupational health and safety laws and regulations The seller shall not be liable for direct indirect consequential incidental or other damages resulting from the sale and use of any goods and sellers liability hereunder shall be limited to repair or replacement of any goods found defective Thi...

Page 5: ...e atmosphere CAUTION This product is intended for both indoor and outdoor use when protected from splashed or wind blown liquids The unit is not waterproof so never submerge the unit in water Pump failure or faulting may result CAUTION This product presents a possible static hazard Do not rub the product with a dry cloth CAUTION Use only the battery pack and chargers specified in this manual Do no...

Page 6: ...nical support contact details In the first instance please contact your supplier If they are unable to resolve your query please contact Markes International on the details below Website www markes com E Mail support markes com Telephone 44 0 1443 230935 UK office 49 0 69 6681089 10 German office 1 866 483 5684 US office toll free 86 21 5465 1216 China office 1 7 Certifications WARNING Intrinsic s...

Page 7: ...A QAR08 0026 05 Notified Body s UL LLC 333 Pfingsten Road Northbrook IL 60062 2096 USA Intrinsically Safe UL File No ALOC7 E118722 US and CAN Class I Zone 0 AEx ia IIC UL 913 Eighth Edition UL 60079 0 Sixth Edition UL 60079 11 Sixth Edition CSA C22 2 No 60079 0 Third Edition CSA C22 2 No 60079 11 14 IEC 60079 0 Sixth Edition IEC 60079 11 Sixth Edition EN 60079 0 2013 RoHS Directive 2011 65 EU Tech...

Page 8: ...attery power is available to run the pump for 8 hours under normal load conditions 5 ON OFF switch 6 Fault indicator This red LED F indicates a flow fault due to excessive back pressure or insufficient battery voltage to maintain flow If flow changes exceed 5 a fault icon appears If a fault exceeds 30 seconds the pump shuts down and the display will show sample time 7 Clock display Shows the conti...

Page 9: ... charger for recharging the internal battery 12 Belt clip 13 Internal rechargeable battery 14 Mode indicator For service use only 15 Mode selector For service use only 16 Hex key For service use only 17 Back pressure adapter Gives the required minimum back pressure the pump needs for low flow TD use Use this in conjunction with the tubing with standard and 6 mm o d TD tubes for flows up to 100 mL ...

Page 10: ...n tube you will not need to re calibrate when switching between tubes Ensure the ACTI VOC is fully charged Attach the PTFE tubing to the pump and push your sorbent tube for calibration into the relevant adapter Figure 2 The sampling arrow on the TD tube should be pointing towards the pump NOTE Use the back pressure adapter when setting flows up to 100 mL min use the PTFE tube adapter when setting ...

Page 11: ...g device e g digital flowmeter part no C FLMTR Turn on the flowmeter and the pump and adjust the flow on the ACTI VOC using the flat bladed screwdriver provided with the ACTI VOC Figure 3 When the desired flow rate has been reached turn off the pump and flowmeter and remove the sorbent tube used for calibration from the adapter The pump is now ready for sampling ...

Page 12: ...as shown Figure 4A ensuring the sampling arrow on the TD tube is pointing towards the pump The tube can be held in position with a penclip part no C CL010 Turn on the pump using the flat bladed screwdriver provided Figure 4B and leave to sample for a given time Be aware that different analytes will have different retention volumes on a given sorbent After the sampling period has passed turn off th...

Page 13: ... Figure 5A ensuring the sampling arrow on the TD tube is pointing towards the pump Turn on the pump using the flat bladed screwdriver provided Figure 5B and leave to sample for a given time Be aware that different analytes will have different retention volumes on a given sorbent After the sampling period has passed turn off the pump Figure 5C and record the sample time This is displayed on the ACT...

Page 14: ...ended periods should be recharged every 1 2 months to avoid complete discharge CAUTION Make certain the charger plug is fully inserted into the port on the battery pack CAUTION Ensure the ACTI VOC is fully charged each time as partial charging can reduce the battery capacity over time CAUTION The charger can be damaged by line transients and overvoltage Although this is not a common problem long t...

Page 15: ...p filter should be changed after approximately six months or 250 hours of operation or when needed CAUTION Failure to change the filter as it becomes dirty will decrease the pump s back pressure capability and performance range To change the pump filter follow the instructions below Blow all dust and debris from around the pump filter Grasp the knurled edge of the pump filter and rotate anticlockw...

Page 16: ...part no SERZ 0285 and a pair of tweezers Use the O ring extraction tool to hook out the O ring Figure 7A taking care not to scratch the groove in which the O ring is seated or to damage the white PTFE insert behind Take the replacement O ring and compress it slightly into an oval shape Figure 7B Pinch with tweezers and insert halfway into the back pressure adapter Use the O ring insertion tool to ...

Page 17: ...ription SERZ 1076 Air inlet pump filter assembly SERZ 1070 Bag sampling accessory SERZ 0351 O Ring extraction tool SERZ 0285 O Ring insertion tool SERLFP 5002 Spare charger SERZ 1097 PTFE tubing and TD tube adapter SERLFP 5003 Back pressure adapter U COV10 O Ring low emission size 10 pk 10 U COV45 O Ring low emission 5 92 mm i d pk 10 Part number Description C FLMTR Digital flowmeter C CPLOK CapLo...

Page 18: ...in Pressure range Back pressure up to 25 H2O Flow control 5 of set point or 3 mL min whichever is greater Battery type Rechargeable nickel metal hydride battery pack Battery output 4 8 V 720 mA h Charging Internal external with adapter Operating temperature 20 C to 45 C 4 F to 113 F Storage temperature 40 C to 45 C 40 F to 113 F Charging temperature 0 C to 45 C 32 F to 113 F ...

Page 19: ...c instant fault shutdown function RFI shielding 8 Contact details For technical support please contact your supplier in the first instance If they are unable to resolve your query please contact Markes International s service department E support markes com T 44 0 1443 230935 W www markes com For an instructional product video please visit chem markes com ACTI VOC Scan the code to watch the video ...

Page 20: ...QUI 1086 L 0181 190521 6 ACTI VOC CapLok MTS 32 and SafeLok are trademarks of Markes International Tedlar is a trademark of DuPont ...