background image

 

 

5.

 

Error Code 

  OUE

r〗

   

- when the weight is > 100% FS + 9d 

-OUE

r〗

   

- when the weight is < - 20d 

HHHHHH

 

- when zero reference is > +2% FS 

LLLLLL

 - when zero reference is < -2% FS 

6.

 

Connections 

POWER SUPPLY 
AC110-220V

(DC9V)

4-20mA CURRENT OUTPUT

RS232/485

LOAD CELL

-

DIGITAL INPUT COM

RELAY OUTPUT COM

 

7.

 

Interface Operation 

Serial format 8-N-1: 8 data bits, No parity bit, 1 stop bit.   

1.

 

Adr=00  

- continuous ASCII reverse data output, if weight is -123.45, the output is 54.321- 

2.

 

Adr=99  

- continuous ASCII regular data output, if weight is -123.45, the output is -123.45 

3.

 

Adr=01-98  - data output using ModBus protocol 

8.

 

Operation 

1.

 

Turn on the indicator, it displays the version number and checks the LED from “0” to “9” 

2.

 

Zero

 

 

When the weight is stable between ±2% FS

press

→0←

zero the display and

→0←

is on 

 

When the 

zero reference 

is > +2% FS, it displays 

HHHHHH

 

 

When the 

zero reference 

is < -2% FS, it displays 

LLLLLL

 

3.

 

Tare

 

 

When the weight is stable,

→T←

is on 

 

When the weight is greater than zero and stable,

〖◣◢〗

  is on 

 

Press

→T←

zero display and

→T←

is on 

4.

 

Clear Tare

 

 

When the weight is stable and the weight is net weight,

→T←

is on 

 

Press

→T←

clears the tare and

→T←

is off 

Summary of Contents for TD 2014 Series

Page 1: ...User Manual TD 2014 Series Weighing Controller Applications Electronic platform scale weighing scale etc ...

Page 2: ... Size 95 48 110 mm Weight 0 5 kg Signal output 4 20 mA analog output Options a Opto isolated RS232 or RS485 b 6 relay outputs with 2 opto isolated inputs that can be used to control the output of relay 1 and 2 2 Model Number TD 2014 0 No interface 1 Optical isolated RS232C 2 Optical isolated RS485 AC AC 110 220V 50 Hz available for XK315A2 6A series only DC DC 9V 1A adapter A Basic model with 4 20...

Page 3: ...ntinuous ASCII regular data output if weight is 123 45 the output is 123 45 3 Adr 01 98 data output using ModBus protocol 8 Operation 1 Turn on the indicator it displays the version number and checks the LED from 0 to 9 2 Zero When the weight is stable between 2 FS press 0 zero the display and 0 is on When the zero reference is 2 FS it displays HHHHHH When the zero reference is 2 FS it displays LL...

Page 4: ... to select filter Press AUtP00 Set zero trace range and zero at start range AUtP 00 no zero set at start zero trace range is 0 4d AUtP 10 zero at start range is 20 FS zero trace range is 0 4d AUtP 20 zero at start range is 20 FS zero trace range is 0 8d AUtP 30 zero at start range is 20 FS zero trace range is 1 2d AUtP 40 zero at start range is 20 FS zero trace range is 1 6d AUtP 50 zero at start ...

Page 5: ...a combinations of T and 0 to change the magnification Press rAtE 1 Set AD sample rate 0 10 times second 1 40 times second 2 250 times second Press Or T rAtE 2 Press a combinations of T and 0 to change the sample rate Press 0 00 Save all parameters 10 Calibrations Operation Display Descriptions Press 000000 Go to Calibrations and Configurations menu and enter password 002003 Press Or T 002003 To en...

Page 6: ... 000000 Go to Set Point menu and enter password 110906 Press Or T 110906 To enter password 110906 Press to move blinking digit to the right Press T to increment blinking digit Press 0 to decrement blinking digit Press A1 Set point 1 for relay 1 Press 000150 Press to access set point 1 Press Or T 000200 Press a combinations of T and 0 to set point A1 to 000200 Press A2 Set point 2 for relay 2 Follo...

Page 7: ...g Press En o5 Set the output of 5 relay Follow step 14 to 16 to change relay output setting Press En o6 Set the output of 6 relay Follow step 14 to 16 to change relay output setting Press 0 00 Save set points 13 Software version VER2 01 packing scale VER2 02 3 speed rating scale VER2 03 check weighing VER2 04 2 materials 2 speed batching scale VER2 05 3 materials 2 speed batching scale VER2 06 4 m...

Page 8: ...7 maintenance and calibration facilities 16 Relay output ...