Servo… User´s manual
English edition
Order No: 65 14 892 E313E
Infant
Adult
Universal
Options
Procedure Diagram: Changing the Patient
Category in Running Mode
4.4.2.2 Selecting the Type of
Ventilation
To change the type of ventilation press
Menu>Biomed>Edit configuration>Start-up
configuration
and follow these steps.
1. Press
Invasive ventilation
OR
2. Press
NIV (Non-Invasive Ventilation)
Note:
The background color on the touch
pads changes when NIV is activated.
3. After confirming the start-up
configuration, press the
Standby
key
when ready to start ventilation.
Note:
Changing the ventilation type affects
the following settings:
• default values for alarm limits
• allowed ranges for alarm limits
• default values for breathing parameters
• allowed ranges for breathing parameters
• pressure and flow regulation
• scaling
Note:
The factory default values may have
been changed by a previous user.
Summary of Contents for SERVO-i
Page 43: ...User s manual VENTILATOR SYSTEM SERVO i V3 0...
Page 44: ......
Page 50: ...vi...
Page 68: ...18 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Page 92: ...42 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Page 102: ...52 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Page 110: ...60 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...