Adding medication
Note: Do not use a syringe with a needle
to add medication.
•
Open the filler cap plug on the nebulizer
unit.
•
Use a pre-filled ampoule or syringe to add
medication into the filler port of the
nebulizer unit.
•
Close the filler cap plug.
Note: Medication can also be added in this
manner during nebulization. This does not
interrupt nebulization or ventilation.
Nebulization
Nebulization can be either:
•
switched on/off for a certain period of time
•
continuous, only Aerogen Solo
Nebulization On/Off
To operate the nebulizer for a specific period
of time (on/off):
1.
Tap MANEUVERS/NEBULIZATION.
2.
Tap Time if more or less than 10 minutes
is required and adjust the figure up or
down.
1
3
2
If the nebulizer cable is not connected, a
dialog Check nebulizer cable connection is
displayed on the screen. Tap OK to accept.
3.
Tap START NEBULIZATION PERIOD. The
default is 10 minutes.
4.
Check that nebulization is in progress and
how much time remains by looking at the
progress symbol.
106
SERVO-air Ventilator System v2.2, User's Manual
| 6 |
Ventilatory settings and functions
|
Summary of Contents for SERVO-air
Page 1: ...x User s Manual SERVO air Ventilator System v2 2...
Page 2: ......
Page 4: ...4 SERVO air Ventilator System v2 2 User s Manual Infologic 1 48 TABLE OF CONTENTS...
Page 16: ...16 SERVO air Ventilator System v2 2 User s Manual 1 Introduction...
Page 84: ...84 SERVO air Ventilator System v2 2 User s Manual 5 Ventilation modes...
Page 110: ...110 SERVO air Ventilator System v2 2 User s Manual 6 Ventilatory settings and functions...
Page 132: ...132 SERVO air Ventilator System v2 2 User s Manual 8 Service Settings...
Page 168: ...168 SERVO air Ventilator System v2 2 User s Manual 10 Index...
Page 169: ...x...
Page 171: ...SERVO air Ventilator System v2 2 User s Manual...