12
1118.01M0
GA1118M1GB01
IV. Operating the operating table
1.Equipotential bonding:
Connect the equipotential bonding pin (
19
) and the equali-
sation wire to an equipotential bonding system in the
operating room.
Î
Note
In countries where potential equalization is required,
the operating table has to be connected to the
potential equalization system of the operating
room. For this purpose, connect the potential
equalization pin of the operating table to one con-
nector of the potential equalization pin of the oper-
ating table to one connector of the potential equali-
zation pin of the operating room.
In countries where potential equalization is not
used, it is not required to connect the operating
table to the potential equalization system.
If no potential equalization cable (5 m long) with
connectors at both ends is included in the scope
of supply of your operating table and you are re-
quired to use PE cable during surgery please or-
der p/n 0200 403 4 at your local MAQUET repre-
sentative.