Infrared Probe 40.01-TX/RX
5
±
30°
Infrared Signal Range / Power Off Delay / Battery Life
Infrared signal range:
There are 3 infrared range settings which can be selected: 1.6m – 3.3m – 4.0m.
A reduced range can be useful to ensure that nearby infrared systems are not interfered with.
A high range setting shortens the battery life.
Power off delay:
If the probe is not triggered within a given time, it is automatically shut off.
3 settings for the delay time are possible: 60s – 120s – 180s.
For removing the shank, see page 6
Attention !
e
After removing the shank, wipe the probe
clean and dry !
Do NOT use compressed air to clean!
e
Ensure the probe is turned off !
e
Adjust the desired range !
e
Remount the shank.
e
Align the stylus !
e
Calibrate the probe !
Maintenance Cleaning
Jets
Dirt can accumulate under the metal eyelid
Cleaning the workpiece and the stylus tip
To clean :
with through-spindle air.
e
Remove service cover with eyelid and
conical spring by hand.
e
Clean probe and parts under flowing water.
e
Replace parts and close probe by hand.
e
Calibrate probe.
Attention !
e
Do NOT use compressed air or high pressure
water
to
clean!
e
Do NOT use any sharp tools!
(These could damage the inner seal)
Optional:
Cleaning jets 40.01-Jets
Infrared Signal
Range
(Typical)
Power Off
Delay
Battery Life in Continuous Use
CR2
5%
usage
½ AA
5%
usage
0
≤
4.0m
Infrared
OFF 180h 130d 270h 190d
1
≤
4.0m
180s
180h 130d 270h 190d
2
≤
3.3m
180s
220h
145d
350h
210d
3
≤
1.6m
180s
290h 170d 440h 240d
4
≤
4.0m
120s
180h 130d 270h 190d
5
≤
3.3m
120s
220h 145d 350h 210d
6
≤
1.6m
120s
290h 170d 440h 240d
7
≤
4.0m
60s
180h 130d 270h 190d
8
≤
3.3m
60s
220h 145d 350h 210d
9
≤
1.6m
60s
290h 170d 440h 240d
Attention !
Only use through-
spindle air !
Pressure < 7bar
Do NOT use coolant !
Screwdriver
Service cover
#3240
Metall eyelid
#2906
Conical spring
#2931
Inner Seal
Factory Settings