www.mc-techgroup.com
SS-5, SS-5/3 | 1.00.01
13
12.2
ELECTRICAL CONNECTIONS
W A R N I N G !
False supply voltage can demage the equipment. When connecting
the equipment, please ensure tha the supply voltage is identical
with the information provided on the model type plate !
N O T E !
For the erection of power installations with rated voltages up to
1000V, the requirements of VDE 0100 and relevant standards and
specifications must be observed!
The main circuit is equipped with a fuse corresponding to the
nominal current (over current protection); for electrical details see
technical data.
The
SS-5...
gas conditioning system is available with either 230V/50 Hz or with 115V/60Hz (for circuit
diagram see Appendix). A 4A fuse is used on all models as fuse protection. The fuse is located on the
clamp mounting rail (see figure 2). In the event that a temperature controller is used in conjunction with
heated sample lines, the overload protection level is increased to 10A.
The electrical connection for the mains is carried out at the terminals 1, 2 and 3 of the mounting rail
(see Fig. 2).
Power in
PE
1 PE
Casing/PE
B2/1 (43)
(15)
7
X1/4
X1/11
X1/2
B2/L1
B1/1
3
2
4
Si
5
6
E1/X2-3
X1/3
8
9
11
10
12
E1/X2-1
Power in
L
Power in
N
E1/X1-1
M2/L
X2/4
M1/blue
M1/black
M1/white
M1/orange
E1/X1-2
B1/3
B1/2
B2/N
(10)
X2/PE
14
13
15
16
17
E1/X1-3
X2/2
M2/N
M2/PE
M1/PE
PE
Connection 115V
X1
with option LA
B1/24
PE
PE
Power in
PE
1 PE
Casing/PE
B2/1 (43)
(15)
7
X1/4
X1/11
X1/2
B2/L1
B1/1
3
2
4
Si
5
6
E1/X2-3
X1/3
8
9
11
10
12
E1/X2-1
Power in
L
Power in
N
E1/X1-1
M2/L
X2/4
M1/blue
M1/black
M1/white
M1/orange
E1/X1-2
B1/3
B1/2
B2/N
(10)
X2/PE
14
13
15
16
17
E1/X1-3
X2/2
M2/N
M2/PE
M1/PE
PE
Connection 230V
X1
with option LA
B1/24
PE
PE
Figure 3
Electrical connections for SS-5 and SS-5/3, 115V and 230V