NOTE!
When analysing a gas mixture, the single components have to be
checked whether there are cross sensitivities, and these have to be con-
sidered during the zero point calibration (see chapter 16.1.1 and 16.1.2).
•
Close the pressure controller output valve and the cylinder valve.
•
Remove the tube connection from the analyzer.
•
The zero point calibration is finished.
NOTE!
After the zero point calibration, also the final value has to be calibrated
again.