Fig. 81: Screen showing details about a single calibration procedure
A screen with detailed information about the failed calibration procedure opens. In this
example it says that the measured value is too high. The measuring range needs to be
adjusted to include the measured value.
Note
Manual calibration with “Zero gas” or “Span gas” can be repeated at
any time.
11.3 Automatic Calibration (AutoCal)
In addition to the manuall calibration (ManuCal) an automatic calibration (AutoCal) feature
is available for single and multi channel multigas analyzers starting from software version
2.24. Follow these steps for automatic calibration. The numbers refer to 'Fig. 81 Screen
showing details about a single calibration procedure':
1
: Activate AutoCal on the M4/S2 System Settings page. Enter the number of
the selected channel for activation.
Fig. 82: Activating AutoCal and entering parameters
1
Activating AutoCal procedure
2
Choose zero or span gas or both
3
Enter starting time for AutoCal procedure
4
Select interval of the AutoCal procedure
5
Enter holding time of the solenoid valves
5
1
4
2
3
Summary of Contents for GenTwo PMA1000V2.4
Page 90: ...89 GenTwo PMA1000V2 4 1 00 00 www mc techgroup com...
Page 91: ...90 GenTwo PMA1000V2 4 1 00 00 www mc techgroup com...
Page 93: ...92 GenTwo PMA1000V2 4 1 00 00 www mc techgroup com...
Page 94: ...93 GenTwo PMA1000V2 4 1 00 00 www mc techgroup com...
Page 95: ...94 GenTwo PMA1000V2 4 1 00 00 www mc techgroup com...
Page 96: ...95 GenTwo PMA1000V2 4 1 00 00 www mc techgroup com...
Page 97: ...96 GenTwo PMA1000V2 4 1 00 00 www mc techgroup com...
Page 98: ...97 GenTwo PMA1000V2 4 1 00 00 www mc techgroup com...
Page 99: ...98 GenTwo PMA1000V2 4 1 00 00 www mc techgroup com...
Page 109: ...108 GenTwo PMA1000V2 4 1 00 00 www mc techgroup com Room for your own notes...