SUPPLY CONNECTIONS
14.1
HOSE/TUBE CONNECTIONS
The gas inlet and outlet hoses/tubes of versions
CG-2/-2M
are connected on the rear of the converter.
Here, the standard executions provides G 1/4" threaded joints. The versions
CG-2H-W/-2MH-W
as wall
mounting execution with heated sample gas inlet and outlet are equipped with a 6 mm tube connector
of stainless steel.
The appropriate tube or hose threaded joint connections are available optionally from M&C.
N O T E !
Do not confuse hose/tube connections for sample gas inlet and out-
let; the connections are labelled accordingly!
Check for tightness after connection!
When connecting the sample gas supply hoses or tubes to the corresponding threaded connections,
pay attention to the following points:
N O T E !
The tightness of the connections can only be guaranteed if the end
section of the connection hose/tube is flat (use a hose-cutter)!
•
Loosen the sleeve nut of the clamping-ring threaded joint by turning to the left. Take care that the nut
is removed carefully from the body of the threaded joint to avoid losing the clamping ring which is
mounted loose in the nut.
•
Push the sleeve nut over the connection hose/tube.
•
Push the clamping ring respectively the pressure and the cutting ring onto the connection hose/tube
with the thicker bulge pointing to nut.
•
Push the hose/tube onto the supporting nipple in the threaded joint.
•
Tighten the sleeve nut by hand.
The hose/tube is now mounted in such a way that it cannot slip and is resistant to pressure.
Connecting a heated sample line with an inner tube of PTFE to the converters
CG-2H-W/-2MH-W,
a
respective insert is absolutely necessary.
Summary of Contents for CG Series
Page 1: ...Gas Converter CG CG 2 2M CG 2H W 2MH W Instruction Manual Version 1 04 00 ...
Page 8: ...8 CG 2 1 04 00 www mc techgroup com Hot surface Contact may cause burn Do not touch ...
Page 33: ...www mc techgroup com CG 2 1 04 00 33 Figure 11 Wiring plan CG 2M Drawing No 2224 5 04 0 ...
Page 34: ...34 CG 2 1 04 00 www mc techgroup com Figure 12 Wiring plan CG 2MH W Drawing No 2224 5 04 4 ...
Page 35: ...www mc techgroup com CG 2 1 04 00 35 Figure 13 Wiring plan CG 2H W Drawing No 2224 5 04 5 ...