Manamed Nature's Bond NB603 Instructions For Use Manual Download Page 14

13

   Immunity Test             IEC 60601 Test Level          Compliance Level          Electromagnetic Environment 

 

 

 

 

 

 

 

                  Guidance

  

Conducted RF
IEC 61000-4-6

Radiated RF
IEC 61000-4-3

3 Vrms
150 Hz to 80 MHz

3 V/m

80 MHz to 2.5 GHz

3 Vrms  

 

 
 

3 V/m

Portable and mobile RF 

communications equipment 

should be used no closer to 
any part of the 

MODEL NB603

including cables, than the 
recommended separation 

distance calculated from the 
equation applicable to the 

frequency of the transmitter.

Recommended separation 
distance

Recommended separation 
distance:

d = 1.2

P

d = 1.2

P 80MHz to 800MHz

d = 2.3

P 800MHz to 2.5GHz

where P is the maximum
output power rating of the 

transmitter in watts (W)
according to the transmitter 
manufacturer and d is the 

recommended separation.

Distance is metres (m)
Field strengths from fixed 
RF transmitters, as 
determined by an 
electromagnetic site survey, 

(a) should be less than the
compliance level in each

frequency range (b).
Interference may occur in 

the vicinity of equipment 
marked with the following 

symbol:

NOTE 1: At 80 MHz, the higher the frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption 

and reflection from structures, objects and people.

a.  Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land 
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with 
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey 
should be considered. If the measured field strength in the location in which the 

MODEL NB603

 is used 

exceeds the applicable RF compliance level above, the 

MODEL NB603

should be observed to verify normal 

operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or 
relocating the 

MODEL NB603

b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. 

The 

MODEL NB603

 is intended for use in the electromagnetic environment specified below. 

The customer or the user of the 

MODEL NB603

 should assure that it is used in such an 

environment.

Guidance and Manufacturer’s Declaration - Electromagnetic Immunity

Summary of Contents for Nature's Bond NB603

Page 1: ...www manamed com 888 508 0712 Breast Pump Instructions For Use Model NB603 IFUNB603 Rev A 10132020...

Page 2: ...ds Recycled Manufacturer Catalog Model Number Protected against solid foreign objects of 12 5mm and greater and vertically falling water drops IP21 ATTENTION Consult ACCOMPANYING DOCUMENTS This symbol...

Page 3: ...perating Cycle At Expression Stage Approx 60 times min AC Input Voltage Range 100 240VAC Operating Frequency 47 63HZ Input Current 600mA Max Output Current 2A Max Output Power 18W DC Output Voltage Ra...

Page 4: ...ssary when used near children or people with disabilities Use product only for its intended use Do not use attachments not recommended by the manufacturer Keep cord away from heated surface Never use...

Page 5: ...the power source Plug the power adaptor into the breast pump first then into the power source Do NOT use antibacterial or abrasive cleaners detergents when cleaning breast pump or breast pump parts N...

Page 6: ...ondbreast pump is intended for lactating women to express and collect milk from their breasts to complement breast feeding Description Nature s Bondis a personal use electric breast pump that has two...

Page 7: ...soft brush 6 Rinse all separated parts with clear water 7 Allow all parts to air dry in a clean area 8 Store dry parts when not in use When using dishwasher 1 Wash all separated parts on top rack of d...

Page 8: ...th no areas of hardness Device Operation Wash hands before touching breast pump kit and breasts Avoid touching the inside of containers or lids Always inspect breast shields connectors valves and tubi...

Page 9: ...ealed bottle in a bowl of warm water for 20 minutes to bring it to body temperature If you are adding expressed breast milk to a container of already frozen breast milk make sure to add a lesser amoun...

Page 10: ...ories to be free from defects in workmanship and materials for a period of ninety 90 days from the date the Accessories are delivered to the original purchaser AccessoriesWarranty Period This LimitedW...

Page 11: ...AMED SHALL NOT BE LIABLE FOR ANY EXEMPLARY OR PUNITIVE DAMAGES Appendix 1 Important information regarding Electro Magnetic Compatibility EMC With the increased number of electronic devices such as PC...

Page 12: ...t 3 Guidance and Manufacturer s Declaration Electromagnetic Emissions The MODEL NB603is intended for use in the electromagnetic environment specified below The customer or the user of the MODEL NB603s...

Page 13: ...5s 3 A m 6 kV contact 8 kV air 2 kV for power supply lines and dc lines 1 kV line s and neutral 2 kV line s to earth 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT...

Page 14: ...he higher the frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a...

Page 15: ...requency of transmitter m output power of transmitter W 0 01 0 1 1 10 100 150 kHz to 80 MHz d 1 2 P 0 12 0 38 1 2 3 8 12 80 MHz to 800 MHz d 1 2 P 0 12 0 38 1 2 3 8 12 800 MHz to 2 5 GHz d 2 3 P 0 23...

Page 16: ...www manamed com 888 508 0712...

Reviews: